Fig. 1.
Schematic diagram of the dilatometer: a) side view; b) top view. 1 – LVDT, 2 – pulley with load, 3 – holder, 4 – carrier alumina rods, 5 – push-rod, 6 – furnace, 7 – heating element, 8 – sample, 9 – supporting rod, 10 – micrometric zeroing screw, 11 – thermocouple.

Fig. 2.
Cross-section of the dilatometric cell. 1 – four carrier rods, 2 – sample, 3 – push-rod or supporting rod.

Fig. 3.
Dilatometric reference line plotted from the tabulated values in [20] (
), the mean value of the repeatedly measured reference sample (orange), correction function with error bars and mean value line (grey), and baseline (green).

Fig. 4.
Comparison of the RLTE of the quartz monocrystal along its c-axis [22],[23],[24],[25],[26] with the experimentally determined RLTE (solid line).

Fig. 5.
Thermodilatometric analysis of a clay sample. The solid line represents RLTE (left axis), the mean of five measurements, with the gray band indicating the error bars. The dotted line is the corresponding CLTE (right axis), obtained as the temperature derivative of the RLTE. The extrema of the CLTE curve identify the temperatures of maximum intensity of the individual processes: release of physically bound water (20 °C → 300 °C), dehydroxylation (400 °C → 750 °C), sintering (above 700 °C), and the formation of new phases (above 900 °C).
Table 1.
The mean value , standard deviation s, and standard uncertainty of the A-type uA(xi).
| at 20 °C | at 800 °C | |||
|---|---|---|---|---|
| k / mm/V | l0 / mm | tm / °C | ΔU / V | |
| −1.907 | 32.23 | 797.6 | 0.025 | |
| s | 0.004 | 0.025 | 2.07 | 0.004 |
| uA(x) | 0.002 | 0.010 | 1.20 | 0.002 |
Table 2.
List of uncertainties of RLTE at room temperature and 800 °C.
| Uncertainty source | Type | Sensitivity coefficient A(x) | Standard uncertainty u(x) | A(x)u(x) [-] | |
|---|---|---|---|---|---|
| 1 | Repeatability of k at 20 °C | A | 7.75 × 10–4 V/mm | 0.002 mm/V | 1.55 × 10–6 |
| 2 | Repeatability of l0 at 20 °C | A | 4.59 × 10–5 1/mm | 0.01 mm | 4.59 × 10–5 |
| 3 | Repeatability of ΔU at 800 °C | A | −0.059 1/V | 0.002 V | −1.18 × 10–4 |
| 4 | Repeatability of tm transformed into Δl at 800 °C | A | 4.59 × 10–5 1/mm | 2×10–4 mm | 9.18 × 10–7 |
| 5 | The sum of A-type uncertainties uA(ɛ), Eq. (9) | ~1.27 × 10–4 | |||
| 6 | Caliper, MAE = 0.01 mm, | B | 4.59 × 10–5 1/mm | 5.77 × 10–3 mm | 2.65 × 10–7 |
| 7 | mV-meter MAE = 1 × 10–5 V, | B | −0.036 1/V | 5.77 × 10–6 V | −2.07 × 10–7 |
| 8 | mV-meter MAE = 1 × 10–5 V, | B | −0.036 1/V | 5.77 × 10–6 V | −2.07 × 10–7 |
| 9 | Micrometer, r = 0.01 mm, | B | 4.59 × 10–5 1/mm | 2.9 × 10–3 mm | 1.33 × 10–7 |
| 10 | Thermocouple at 800 °C | B | 4.59 × 10–5 1/mm | 8.0 × 10–3 mm | 3.67 × 10–7 |
| 11 | Temperature controller at 800 °C | B | 4.59 × 10–5 1/mm | 1.8 × 10–3 mm | 1.00 × 10–7 |
| 12 | Cold junctions | B | 4.59 × 10–5 1/mm | 3.0 × 10–4 mm | 1.37 × 10–8 |
| 13 | Inhomogeneity of temperature field at 800 °C | B | 4.59 × 10–5 1/mm | 6.0 × 10–4 mm | 2.75 × 10–8 |
| 14 | Pressure on sample | B | 4.59 × 10–5 1/mm | 2.5 × 10–5 mm | 1.15 × 10–9 |
| 15 | Sample position | B | 4.59 × 10–5 1/mm | 2.0 × 10–3 mm | 9.21 × 10–8 |
| 16 | The sum of B-type uncertainties uB (ɛ), Eq. (11) | 5.72 × 10–7 | |||
| 17 | The combined uncertainty uC (ɛ), Eq. (12) | ~1.27 × 10–4 | |||
| 18 | The final value of RLTE: ɛ = 0.0038 ± 0.0002 | ||||
Table 3.
List of uncertainties of CLTE at 800 °C.
| Uncertainty source | Type | Sensitivity coefficient A(x) | Standard uncertainty u(x) | A(x)u(x) [-] | |
|---|---|---|---|---|---|
| 1 | Temperature step Δt | B | 6.08 × 10–4 1/K2 | ~0.1 K | 6.08 × 10–5 |
| 2 | Linear thermal expansion (Δl/l0)s | B | 0.4 1/K | 1.27 × 10–4 - | 4 × 10–5 |
| 3 | The combined uncertainty uc (α) | 7.28 × 10–5 | |||
| 4 | The final value of CLTE: α = (1.52 ± 0.14) × 10–3 1/K | ||||