1. Introduction
Solid materials, such as ceramics and their constituent minerals, exhibit thermal expansion upon heating. In specific instances, structural modifications, such as dehydroxylation in natural clays, occur during heating. These transformations are typically accompanied by significant variations in specific volume, which can be measured by the corresponding change in sample length. The measurement of its thermal dilatation during linear heating/cooling or isothermal heating, called thermodilatometric analysis (TDA), is an important tool for determining phase-transformation kinetics, sintering in ceramics, optimizing the firing regime, estimating mineral composition, and other applications [1],[2],[3],[4],[5].
The physical quantity characterizing thermal expansion is the coefficient of linear thermal expansion α (CLTE) derived from the relative linear thermal expansion (RLTE). Let a sample have an initial length l0 at room temperature t0. When the sample is heated, it expands to l1 at t1, and then to l2 at t2. The instantaneous CLTE within the temperature interval (t1, t2) at the mean temperature tm is defined as [6]
The derivative defines the instantaneous CLTE dl/dt at a single temperature. This represents the limiting case of (1) as t1 and t2 become very close to each other. The fundamental basis for determining the CLTE is the RLTE Δl/l0; its derivative d(Δl/l0)/dt yields the CLTE.
A survey of dilatometric results obtained for fired ceramics indicates that Δl/l0 < 1 % within the temperature interval of 20-1300 °C. During the sintering of the green sample, a significant contraction Δl/l0 > 12 % can be observed in this temperature interval. In addition, certain phase changes are accompanied by large shrinkage, e.g., the intensive rebuilding of the crystal lattice during dehydroxylation and the formation of new phases at 950 °C in clay ceramics. Consequently, a large measurement range, rather than extreme sensitivity, is required for TDA of green clay ceramics. A resolution of 1 × 10–6 m is sufficient for most experimental investigations. A push-rod dilatometer with a linear variable differential transformer (LVDT) and alumina support elements (tube or rods) effectively fulfills these requirements.
A differential horizontal alumina dilatometer, designed and constructed in the authors' laboratory, was used to measure the thermal expansion of kaolinite- and illite-based samples, as well as ceramic components (quartz, feldspar, calcite, kaolin, and various clays) during heating from room temperature to 1250 °C [5],[6],[7],[8],[9],[10],[11].
The calibration of relative dilatometers is infrequently addressed in the research literature. Furthermore, the authors have not encountered studies that focus on uncertainty analysis of TDA results obtained during thermal treatment. The objective of the present study is to provide a detailed description of the dilatometer calibration and to estimate the measurement uncertainty associated with RLTE and CLTE.
2. Dilatometer
The push-rod dilatometer uses a differential measurement technique to determine the relative displacement between the expansion of alumina carrier components (rods or tubes) with a known RLTE and that of a sample with unknown thermal expansion.
Different techniques for measuring the thermal expansion of solids are briefly reviewed in [1], [12],[13],[14]; however, no previously reported dilatometric cell design is identical to the configuration used in the present study. A similar construction was reported in [15], [16]. A dilatometric cell, as illustrated in Fig. 1 and Fig. 2, consists of four carrier alumina rods (⌀3.0 mm) secured to two duralumin holders, along with the push-rod and supporting rod.

Fig. 1.
Schematic diagram of the dilatometer: a) side view; b) top view. 1 – LVDT, 2 – pulley with load, 3 – holder, 4 – carrier alumina rods, 5 – push-rod, 6 – furnace, 7 – heating element, 8 – sample, 9 – supporting rod, 10 – micrometric zeroing screw, 11 – thermocouple.

Fig. 2.
Cross-section of the dilatometric cell. 1 – four carrier rods, 2 – sample, 3 – push-rod or supporting rod.
The LVDT (INPOS, ZPA Jinonice, Czech Republic) is fixed to one of the holders. The assembly features a lightweight design, with neither holder being rigidly fixed to the apparatus frame. The cell assembly passes freely through the furnace openings, which are insulated with ceramic fiber.
The push-rod is pressed against the sample with a weight to enable loading for dilatometry (e.g., compressive creep). The furnace is constructed from porous alumina bricks, and four silicon carbide (SiC) rods are used as heating elements. The operational temperature range of the dilatometer is 20 °C to 1250 °C under a static air atmosphere.
The central part of the dilatometric cell, containing the sample, is located within the furnace. A cross-sectional view of the dilatometric cell illustrating the spatial arrangement of the carrier rods, sample, and push-rod is presented in Fig. 2. The sample is placed between two rods: the push-rod and the stationary supporting rod, the latter being fixed to the second holder. The axial position of the push-rod, specimen, and supporting rod assembly is adjustable via a micrometric screw. The LVDT measures the displacement of the push-rod, Δd, which corresponds to the sample's relative longitudinal deformation. An Agilent 34972A DC multimeter measures analog voltage signals from both the LVDT and the thermocouple, then digitizes and stores them on a personal computer. Temperature expansion measurements are performed in adjustable steps ranging from 10 s to 3 min.
The furnace temperature is measured using a Pt-PtRh10 thermocouple placed near the sample. The temperature is controlled by a temperature programmer, Clasic Clare 4 (Czech Republic), operating independently of the PC. The multimeter performs internal cold junction compensation and voltage-to-temperature conversion for the Pt-PtRh10 thermocouple.
The presented design allows adjustment of the measurement cell for samples with dimensions ⌀5 × 25 mm up to ⌀10 × 70 mm. This flexibility enables the measurement of inhomogeneous materials that contain coarse grains or textured microstructure. The constant pressure of the push-rod on the sample is achieved by gravity acting on the weight. This further permits adjustment of the pressing force and monitoring of compression creep behavior.
The basic scheme of the dilatometer shown in Fig. 1 highlights that only differential expansion in section B contributes to the LVDT signal, as expansions in sections A and C are compensated. The LVDT measures ΔU = U − U0 voltage at elevated temperature t, resulting from the relative displacement of the push-rod, Δl, between the sample and alumina carrier rods, which expand in opposite directions. The initial voltage U0 is adjusted at room temperature, assuming the linear relationship between Δl and ΔU, i.e. Δl = kΔU. The following expression holds for an ideal dilatometer:
where Δls is the change in the length of the sample, Δlar is the change in the length of the carrier alumina rods at temperature t, and k is the experimentally determined LVDT constant.The relative expansion of the alumina rod is known across the entire temperature range. By defining l0 as the initial sample length, the RLTE of the sample is calculated using the known RLTE values of alumina Δlar/l0 and the measured values of ΔU. The RLTE of the sample Δls/l0 is then derived from (2) as follows:
In (3), a term Δlse/l0 is incorporated to account for the contribution of systematic errors (e.g., zero drift) inherent to the real dilatometer. Once the RLTE of the sample is determined, CLTE over the temperature interval Δt can be calculated as
3. Calibration
Since the employed method is not absolute, the calibration of the dilatometer is necessary [6], [17],[18],[19]. Calibration was performed at a heating rate of 5 °C/min, a commonly used thermal regime in the authors' laboratory. For this purpose, a reference sample for dilatometry (⌀5 × 30.00 mm) of X10NiCrMoTiB 1515 steel, provided by the Physikalisch-Technische Bundesanstalt (National Metrology Institute), Berlin, was used. This material has been intensively studied in recent studies. Tabulated data from the round-robin test [20] were adopted as the reference dilatometric curve (Fig. 3), which the following linear function can express
where t is in °C and ɛref is in %.
Fig. 3.
Dilatometric reference line plotted from the tabulated values in [20] (
), the mean value of the repeatedly measured reference sample (orange), correction function with error bars and mean value line (grey), and baseline (green).
As previously noted, the carrier rods compensate for the thermal expansion of the push-rod and the support rod when their temperatures are equal. In a real dilatometer, an additional small displacement of the push-rod arises from minor temperature differences among the dilatometer elements during heating. This undesirable displacement can be characterized by performing a TDA test using a sample from the alumina rod. The resulting data define the dilatometer's baseline for the specific thermal regime. The mean gradient of the baseline (Fig. 3) of the calibrated dilatometer was 2.62 × 10–7 1/°C, which was below the maximum permissible limit of 3 × 10–7 1/°C for a properly configured dilatometer [6]. To eliminate this error, the calibration using the reference sample must be performed [17].
The first stage of the calibration procedure involved the experimental determination of the LVDT constant, k, in (2) at room temperature. The value of k was obtained by measuring the output of the LVDT ΔU = U − U0 (U0 denotes the initial voltage) and the displacement Δl of the push-rod controlled by a micrometric screw. The obtained LVDT constant k = −1.907 ± 0.004 mm/V was used in subsequent analysis.
The reference sample was measured 5 times at a heating rate of 5 °C/min. The experimental data obtained over the temperature range 20 °C – 1000 °C (Fig. 3) were used for calibration. The difference Δɛ(t) between the reference data ɛref (t) and the experimental data ɛexp (t) must be added to the experimental data to obtain the reference values, i.e., ɛref (t) = ɛexp(t) + Δɛ(t). This difference, Δɛ(t), is the correction function and corresponds to the term in parentheses in (3). The analytical expression of the correction function obtained via polynomial fitting is defined as
where t is in °C and ɛcor is in %.Equation (3) can be written using (6) as
where ɛ is in %, l0 in mm, ΔU = (U − U0) in V, tm = (t2 + t1)/2 in °C, and k = −1.907 mm/V. The temperatures t1 and t2 represent the most proximal values centered around ~800 °C.In addition to mechanical calibration of the dilatometer, a temperature calibration should be performed if a high-precision TDA is required. The procedure involves heating a composite specimen consisting of thin metallic foils sandwiched between alumina segments. Upon reaching the melting point of the metallic standard, a distinct discontinuity is observed in the TDA curve. This technique enables the determination of the deviation between the actual specimen temperature and the recorded temperature [18], [21]. Given that the kinetic processes in ceramic specimens typically proceed gradually over extended temperature intervals, extreme temperature precision is not strictly required for these applications. A simple test based on the melting point of aluminum (660.3 °C), the α → β transition of quartz (573 °C), the α → β transition of cristobalite (~220 °C), and the metakaolinite lattice collapse (~950 °C) confirmed that these transformations occur in the dilatometer close to their theoretical characteristic temperatures.
The sensitivity of the dilatometer, as determined by the minimum resolvable increment in the Δls record, is 6 × 10–8 m. The minimal resolvable increment Δls was determined from the smallest detectable change in the output voltage ΔU of the LVDT sensor, according to the relation Δls = kΔU, where k is the calibration constant of the sensor. This determination was performed at a slow heating rate of 2 °C/min to ensure stable signal conditions.
The correctness of the calibration was verified by TDA on a quartz monocrystal along its c-axis at a heating rate of 5 °C/min. The experimental results show very good agreement with the data reported in [22],[23],[24],[25],[26] from five independent sources, see Fig. 4.

Fig. 4.
Comparison of the RLTE of the quartz monocrystal along its c-axis [22],[23],[24],[25],[26] with the experimentally determined RLTE (solid line).
4. Uncertainty analysis of RLTE and CLTE
The standard wet-ceramic processing was used to prepare the samples: mineral powders were mixed with water to form a plastic ceramic mass. A plastic ceramic mixture consisting of kaolinite (50 wt.%), ground quartz (25 wt.%), and albite (25 wt.%), with particle sizes below 200 μm, was used to produce an extruded die-cast rod of ⌀8 × 200 mm. From this rod, samples of ⌀8 × 33 mm were cut, ensuring an identical microstructure across all specimens. After drying, the samples were subjected to TDA at a heating rate of 5 °C/min.
The data in Fig. 5 indicate the temporary scatter (650 °C – 1000 °C). Above 1000 °C, all RLTE curves converge again, and only minimal variation is observed. The temperature range in which the scatter appears corresponds to pronounced microstructural and structural changes caused by solid-state and liquid-phase sintering, as well as the collapse of metakaolinite. These processes inherently involve thermally driven structural rearrangements, making thermal randomness the most probable source of the observed scatter.

Fig. 5.
Thermodilatometric analysis of a clay sample. The solid line represents RLTE (left axis), the mean of five measurements, with the gray band indicating the error bars. The dotted line is the corresponding CLTE (right axis), obtained as the temperature derivative of the RLTE. The extrema of the CLTE curve identify the temperatures of maximum intensity of the individual processes: release of physically bound water (20 °C → 300 °C), dehydroxylation (400 °C → 750 °C), sintering (above 700 °C), and the formation of new phases (above 900 °C).
The uncertainty analysis was performed in two stages: a) for the RLTE of the unfired sample, b) for CLTE. Upon inspecting the TDA curve (Fig. 5), significant scatter in the measured values was observed at 800 °C; therefore, the uncertainty analysis focused on this temperature.
According to [27], measurement uncertainty comprises components of two types, A and B. A-type uncertainties are evaluated through statistical analysis; the uncertainty of the measured quantity xi is defined as
where si denotes the standard deviation, n is the number of independent measurements, and is the mean value.Each dilatometric analysis assumes measuring the initial sample length l0 at room temperature. According to [28], the length l0 should be measured with a relative error Δl/l0 < 0.2 %. The length of the sample l0 was measured six times at room temperature, and the standard uncertainty of l0 is 0.01 mm (see Table 1), or 3 × 10–4 (0.03 %), i.e., the requirement of [28] is fulfilled.
Table 1.
The mean value , standard deviation s, and standard uncertainty of the A-type uA(xi).
| at 20 °C | at 800 °C | |||
|---|---|---|---|---|
| k / mm/V | l0 / mm | tm / °C | ΔU / V | |
| −1.907 | 32.23 | 797.6 | 0.025 | |
| s | 0.004 | 0.025 | 2.07 | 0.004 |
| uA(x) | 0.002 | 0.010 | 1.20 | 0.002 |
The sample used for the uncertainty analysis consisted of illitic clay. Upon heating, the clay sample undergoes irreversible processes, transforming the raw material into a sintered ceramic body. To determine Type A uncertainty, which is derived from repeated measurements, a series of identical samples must be used. Therefore, TDA of the clay sample was repeated 5 times, always with a fresh sample, see Fig. 5. The LVDT output voltage, U, at temperature t (closest to 800 °C) was recorded every 30 s, i.e., in Δt = 2.5 °C steps. The voltages U and U0, as well as temperature t, were extracted from the protocol stored in the PC for each individual TDA.
The mean values of l0, t, and ΔU, with their standard deviations and A-type standard uncertainties uA(x) are presented in Table 1.
The sum of the A-type uncertainties, which follows from the directly measured quantities in (7), is
where sensitivity coefficients A(xi) are partial derivatives of (7)The mean values from Table 1 were substituted into the sensitivity coefficients to obtain their numerical values (see Table 2).
Table 2.
List of uncertainties of RLTE at room temperature and 800 °C.
| Uncertainty source | Type | Sensitivity coefficient A(x) | Standard uncertainty u(x) | A(x)u(x) [-] | |
|---|---|---|---|---|---|
| 1 | Repeatability of k at 20 °C | A | 7.75 × 10–4 V/mm | 0.002 mm/V | 1.55 × 10–6 |
| 2 | Repeatability of l0 at 20 °C | A | 4.59 × 10–5 1/mm | 0.01 mm | 4.59 × 10–5 |
| 3 | Repeatability of ΔU at 800 °C | A | −0.059 1/V | 0.002 V | −1.18 × 10–4 |
| 4 | Repeatability of tm transformed into Δl at 800 °C | A | 4.59 × 10–5 1/mm | 2×10–4 mm | 9.18 × 10–7 |
| 5 | The sum of A-type uncertainties uA(ɛ), Eq. (9) | ~1.27 × 10–4 | |||
| 6 | Caliper, MAE = 0.01 mm, | B | 4.59 × 10–5 1/mm | 5.77 × 10–3 mm | 2.65 × 10–7 |
| 7 | mV-meter MAE = 1 × 10–5 V, | B | −0.036 1/V | 5.77 × 10–6 V | −2.07 × 10–7 |
| 8 | mV-meter MAE = 1 × 10–5 V, | B | −0.036 1/V | 5.77 × 10–6 V | −2.07 × 10–7 |
| 9 | Micrometer, r = 0.01 mm, | B | 4.59 × 10–5 1/mm | 2.9 × 10–3 mm | 1.33 × 10–7 |
| 10 | Thermocouple at 800 °C | B | 4.59 × 10–5 1/mm | 8.0 × 10–3 mm | 3.67 × 10–7 |
| 11 | Temperature controller at 800 °C | B | 4.59 × 10–5 1/mm | 1.8 × 10–3 mm | 1.00 × 10–7 |
| 12 | Cold junctions | B | 4.59 × 10–5 1/mm | 3.0 × 10–4 mm | 1.37 × 10–8 |
| 13 | Inhomogeneity of temperature field at 800 °C | B | 4.59 × 10–5 1/mm | 6.0 × 10–4 mm | 2.75 × 10–8 |
| 14 | Pressure on sample | B | 4.59 × 10–5 1/mm | 2.5 × 10–5 mm | 1.15 × 10–9 |
| 15 | Sample position | B | 4.59 × 10–5 1/mm | 2.0 × 10–3 mm | 9.21 × 10–8 |
| 16 | The sum of B-type uncertainties uB (ɛ), Eq. (11) | 5.72 × 10–7 | |||
| 17 | The combined uncertainty uC (ɛ), Eq. (12) | ~1.27 × 10–4 | |||
| 18 | The final value of RLTE: ɛ = 0.0038 ± 0.0002 | ||||
The CLTE can be obtained via numerical differentiation of RLTE by dividing ɛ by a small temperature interval Δt, see (4). The sensitivity coefficients for CLTE are
The ɛ for 800 °C is the mean value of the RLTE, and Δt = 2.5 °C was determined from the applied heating rate (5 °C/min) and the selected sampling interval (0.5 min).
The B-type uncertainties, uB (xi ), are evaluated from the manufacturer's specifications, calibration certificates, and expert knowledge. These uncertainties originate from various independent sources and represent systematic effects that cannot be treated statistically. For the digital and analog measuring devices used in this study, the B-type uncertainties are expressed as and , where MAE is the maximum admissible error and r is the resolution of the instrument. Additional B-type uncertainties result from factors influencing the displacement of the push-rod, and each of these contributions must be evaluated individually.
The sum of the B-type uncertainties is given by the equation
where m is the number of sources of B-type uncertainties.A. Sources of B-type uncertainties:
The LVDT constant
Using the constant k, the output voltage from the LVDT is converted into push-rod displacement: Δls = kΔU. The LVDT was calibrated using an analog micrometer with a resolution r = 0.01 mm; consequently, the standard uncertainty of the micrometer is , which is equal to uB (k), see line 9 in Table 2.
The sample position
The sample location within the dilatometric cell is strictly defined by its construction; consequently, it does not influence the value of Δls/l0. The sample, push-rod, and supporting rod may not be perfectly centered, which could introduce an estimated shift of 2 × 10–3 mm.
Mechanical pressure on the sample
Reliable mechanical contact between the push-rod, supporting rod, and the sample is maintained by a weight. As a result, a compressive force causes their deformation. Taking into account a compressive force F = 0.1 N, the dimensions of the rods and ceramic sample, as well as the temperature dependence of their Young's moduli, the total deformation is estimated at ~2.5 × 10–5 mm at temperatures below 1000 °C.
The thermocouple aging and corrosion
Exposed Pt-PtRh10 (type S) thermocouples undergo corrosion and aging during service. The aging is minimal at temperatures below 1400 °C, and the influence of corrosion (due to silicon) is low for wires thicker than 0.3 mm. Consequently, the thermocouple was considered a stable sensor with invariant properties.
Note: in the subsequent sections, the estimated temperature deviation Δt was converted into the push-rod displacement Δl for l0 = 32.23 mm and a CLTE of the sample material α = 3.12 × 10−6 1/K around 800 °C:
The thermocouple and temperature control
A typical accuracy of a Pt-PtRh10 thermocouple is within ~1 % of the measured temperature, i.e., ~8.0 °C at 800 °C. This temperature deviation induces a push-rod displacement of ~8.0 × 10–3 mm due to the sample's thermal expansion.
The temperature controller Clasic Clare 4 (Czech Republic) is used to regulate the thermal regime (heating, cooling, or isothermal) within the furnace. According to the manufacturer's specifications, the temperature error is ~1 °C at 800 °C, and the long-term stability is 0.3 °C when the room temperature changes by 10 °C. Consequently, the temperature controller can introduce additional sample expansion ~1.8 × 10–3 mm.
The compensation leads connect the thermocouple wires to the input of an Agilent 34972A multimeter, which performs cold-junction compensation and converts the thermoelectric voltage into temperature. The error associated with cold junction compensation is negligible, at less than 0.3 °C. Combined with the thermocouple error, the cumulative temperature error is ~1.5 °C. This can cause a small shift in the push-rod of ~3 × 10–4 mm.
The influence of temperature field inhomogeneity
When the temperature is not uniformly distributed throughout the sample volume, the sample's properties depend on position. Consequently, a homogeneous temperature field is required in the region surrounding the sample. Since the carrier rods do not shield the sample from the heating elements, the measuring cell temperature is distributed uniformly. The internal dimensions of the furnace are 100 × 70 × 70 mm. The thermal field in the central region of the dilatometric cell was measured at constant temperatures of 900 °C and 1150 °C. It was found that temperature deviations between the sample center and its ends were less than 3 °C at 1150 °C, which falls within the permitted interval specified by [28]. For Δt = 3 °C, the uncertainty is 6.0 × 10–4 mm.
The combined uncertainty of the linear thermal expansion is defined as the square root of the sum of the squared uncertainty components listed in Table 2, lines 6 and 17
The expanded uncertainty is the combined standard uncertainty multiplied by a coverage factor, typically 2. This implies that the true value is presumably within the interval with 95 % probability. The final result of the uncertainty analysis for RLTE is expressed as
where is the mean value derived from (7) and the mean values from Table 1. Upon substituting these numerical values, , i.e., the relative expanded uncertainty of RLTE is ~5.26 %.Temperature measurements and the determination of Δt are conducted under consistent experimental conditions throughout the analysis. Consequently, this uncertainty source is invariant; it depends solely on the dilatometer properties and is independent of the sample properties. The value of Δt was transformed into an equivalent push-rod displacement Δl (see Table 2, line 5). The most significant uncertainty component arises from the scatter of ΔU values measured across a group of nominally identical samples. The scatter of ΔU, which is proportional to the scatter of RLTE, varies with the temperature, as shown in Fig. 5. Consequently, the uncertainty of RLTE is also temperature-dependent. Its values increase for t > 550 °C, where new phases emerge: meta-illite, metakaolinite, β-quartz, glass, and mullite are formed on the background of the sintering. The authors observed a similar effect when analyzing elastic modulus or electrical conductivity [29], [30].
The CLTE value at 800 °C was calculated from (4) using , and Δt = 2.5 °C yields α = 1.52 × 10–3 1/K. The sensitivity coefficients for CLTE, see (10), follow from (4). Their numerical values for Δt = 2.5 °C and are given in Table 3. The relative uncertainty of CLTE is 9.2 % at 800 °C, representing the maximum scatter of the experimental data.
Table 3.
List of uncertainties of CLTE at 800 °C.
| Uncertainty source | Type | Sensitivity coefficient A(x) | Standard uncertainty u(x) | A(x)u(x) [-] | |
|---|---|---|---|---|---|
| 1 | Temperature step Δt | B | 6.08 × 10–4 1/K2 | ~0.1 K | 6.08 × 10–5 |
| 2 | Linear thermal expansion (Δl/l0)s | B | 0.4 1/K | 1.27 × 10–4 - | 4 × 10–5 |
| 3 | The combined uncertainty uc (α) | 7.28 × 10–5 | |||
| 4 | The final value of CLTE: α = (1.52 ± 0.14) × 10–3 1/K | ||||
When the uncertainty analysis is performed at a temperature where only ordinary thermal expansion occurs (e.g., 300 °C), the data scatter is substantially smaller (see Fig. 5). In such cases, the uncertainty of the RLTE is ~1 %, and the uncertainty of the CLTE is ~3 %.
5. Conclusions
The RLTE and CLTE of a kaolinite-based clay sample were measured using a horizontal alumina push-rod dilatometer. The dilatometer is described with respect to possible sources of uncertainty. The uncertainty analysis takes into account the homogeneity of the temperature field, the sample position, mechanical pressure on the sample, thermocouple aging and corrosion, thermocouple and temperature control, and the measurement of the LVDT output. Uncertainties associated with the measurement apparatus are also incorporated. The summation of all partial uncertainty components yields the expanded uncertainty of the thermodilatometric analysis. The uncertainty is temperature-dependent, and the expanded uncertainty reaches the highest value, ~5.26 %, at 800 °C, where transformations in sample properties generate significant data scatter. When only ordinary thermal expansion occurs, the uncertainty of the RLTE is ~1 %, and the uncertainty of the CLTE is ~3 %.