Skip to main content
Have a personal or library account? Click to login
Study of a 2 kN·m Torque Transducer Tested at GUM and PTB, Including Creep Behaviour Cover

Study of a 2 kN·m Torque Transducer Tested at GUM and PTB, Including Creep Behaviour

Open Access
|Apr 2024

Abstract

This article presents a study carried out on a 2 kN·m torque transducer at the Central Office of Measures (GUM) and the Physikalisch-Technische Bundesanstalt (PTB). The weighted least squares method was used to generate the linear regression equations for this torque transducer. The Monte Carlo method and the law of uncertainty propagation were used to calculate the expanded uncertainty. In addition, a creep study was carried out at eight measurement points ranging from 200 N·m to 2000 N·m. The investigations showed that the highest readings of the torque transducer, expressed in electrical units as mV/V, occur within the initial few seconds of the test after the removal of the maximum reference torque.

Language: English
Page range: 88 - 94
Submitted on: Aug 1, 2023
Accepted on: Mar 12, 2024
Published on: Apr 13, 2024
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2024 Janusz D. Fidelus, Anna Trych-Wildner, Jacek Puchalski, Paula Weidinger, published by Slovak Academy of Sciences, Institute of Measurement Science