Have a personal or library account? Click to login
Stability Studies of Torque Transducers Cover
By: Anil Kumar and  Harish Kumar  
Open Access
|Jun 2011

Abstract

The present paper discusses the stability studies carried out on a small number of torque transducers over the years with capacities ranging from 10 Nm to 1000 Nm. The torque transducers have been calibrated using the torque standard machine based on the written standard calibration procedure. The study reveals that the uncertainty of measurement of torque transducers has been varying appreciably and it is more for lower range. Besides, the deviation from their average values has also been studied and found to be varying through the years.

Language: English
Page range: 41 - 44
Published on: Jun 3, 2011
Published by: Slovak Academy of Sciences, Institute of Measurement Science
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2011 Anil Kumar, Harish Kumar, published by Slovak Academy of Sciences, Institute of Measurement Science
This work is licensed under the Creative Commons License.

Volume 11 (2011): Issue 2 (April 2011)