Abstract
Water vapour density inside the mass comparator enclosure is a critical parameter whose fluctuations during mass weighing can lead to errors in the determination of an unknown mass. To monitor them, a method using DFB laser diode in the near infrared has been proposed and tested. Preliminary results of our observation of water vapour sorption and de-sorption processes from the walls and the mass standard are reported.
DOI: https://doi.org/10.2478/v10048-008-0006-y | Journal eISSN: 1335-8871
Language: English
Page range: 22 - 24
Published on: Mar 7, 2008
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open
Keywords:
Related subjects:
© 2008 N. Khélifa, published by Slovak Academy of Sciences, Institute of Measurement Science