Contribution of the SPRT Calibration to Uncertainty of Temperature T90 Measured by the Calibrated SPRT Cover

Contribution of the SPRT Calibration to Uncertainty of Temperature T90 Measured by the Calibrated SPRT

Open Access
|Mar 2008

Abstract

The propagation of uncertainties, when the International Temperature Scale of 1990 (ITS-90) is used by a standard platinum resistance thermometer (SPRT) calibrated at defining fixed points (DFP), can be solved by applying several approaches. The article presents an analysis of contribution of covariance between resistances of SPRT at the defining fixed points (DFP). Its effect on temperature measured by calibrated SPRT is demonstrated by using real calibration data.

Language: English
Page range: 5 - 10
Published on: Mar 7, 2008
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2008 S. Ďuriš, R. Palenčár, J. Ranostaj, published by Slovak Academy of Sciences, Institute of Measurement Science
This work is licensed under the Creative Commons License.

Volume 8 (2008): Issue 1 (February 2008)