Have a personal or library account? Click to login
An Experimental Setup for Power Loss Measurement up to 1 kHz using an Epstein Frame at CMI Cover

An Experimental Setup for Power Loss Measurement up to 1 kHz using an Epstein Frame at CMI

By: Michal Ulvr  
Open Access
|Nov 2023

Abstract

This paper describes an experimental setup used at the Czech Metrology Institute (CMI) to measure the specific power loss of oriented and non-oriented electrical steel sheets up to 1 kHz using an Epstein frame. Special attention is given to a) a description of the hardware that is used, b) a description of the feedback control and measurement software, and c) an analysis of the sources of uncertainty and validation. Calibration expanded uncertainty of (0.5 up to 1.6)% for k = 2 can be achieved with this setup.

Language: English
Page range: 275 - 280
Submitted on: Jul 13, 2023
Accepted on: Oct 27, 2023
Published on: Nov 17, 2023
Published by: Slovak Academy of Sciences, Institute of Measurement Science
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2023 Michal Ulvr, published by Slovak Academy of Sciences, Institute of Measurement Science