Have a personal or library account? Click to login
Development of a Temperature and Heat Flux Measurement System Based on Microcontroller and its Application in Ophthalmology Cover

Development of a Temperature and Heat Flux Measurement System Based on Microcontroller and its Application in Ophthalmology

Open Access
|Mar 2022

Abstract

The paper describes the design and technical parameters of a medical thermoelectric device developed for diagnosing and monitoring the ophthalmic diseases. The main elements of the device are a specially designed thermoelectric heat flux sensor and a thermocouple temperature sensor connected to a data acquisition unit. The sensor is a thermoelectric micro-module that converts the heat flux into an electric voltage, which is recorded by the measuring channel of the data acquisition unit. The device allows high-precision measurements of both heat flux and temperature from the ocular surface. The paper contains examples of clinical piloting of the device.

Language: English
Page range: 73 - 79
Submitted on: Nov 9, 2021
Accepted on: Jan 28, 2022
Published on: Mar 12, 2022
Published by: Slovak Academy of Sciences, Institute of Measurement Science
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2022 Chunzhi Wang, Hongzhe Jiao, Lukyan Anatychuk, Nataliya Pasyechnikova, Volodymyr Naumenko, Oleg Zadorozhnyy, Lyudmyla Vikhor, Roman Kobylianskyi, Roman Fedoriv, Orest Kochan, published by Slovak Academy of Sciences, Institute of Measurement Science
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.