Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements Cover

Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements

Open Access
|Nov 2026
Read online

Abstract

Phase angles measured at low frequencies with the tetrapolar technique are sometimes scattered or even positive. We used Monte Carlo simulations of lock-in detection to quantify how large phase errors that measurement noise alone can cause. For a signal-to-noise ratio after demodulation (SNRout) of about 3 or higher, the standard deviation of the phase error was close to 57° divided by SNRout, and 95% of the errors were within about twice this value. At SNRout = 10 the error standard deviation was 6°; at SNRout = 1 it was 64° and 16% of the measurements were wrong by more than 90°. As SNRout decreased towards zero, the phase gradually approached a completely random value. In an illustrative low-frequency tetrapolar measurement, the absolute phase errors were small (standard deviation 0.007° at 1 Hz and 0.04° at 0.1 Hz), but the true tissue phase was even smaller. The error was therefore 1.2 times the true phase at 1 Hz and 37 times at 0.1 Hz, and the measured phase had the wrong sign in 20% and 49% of the measurements, respectively. The sign of a small phase angle is only reliable (wrong in less than 1% of measurements) when the true phase is larger than 2.3 times the error standard deviation.

Language: English
Page range: 178 - 182
Submitted on: Oct 1, 2026
Accepted on: Oct 29, 2026
Published on: Nov 4, 2026
Published by: University of Oslo
In partnership with: Paradigm Publishing Services

© 2026 Jie Hou, Sisay Mebre Abie, Ørjan G. Martinsen, published by University of Oslo
This work is licensed under the Creative Commons Attribution 4.0 License.