Validation of microwave-assisted sample extraction-high performance liquid chromatography (HPLC) method for quantitative analysis of hydrolysed amino acids in aquatic biological samples Cover

Validation of microwave-assisted sample extraction-high performance liquid chromatography (HPLC) method for quantitative analysis of hydrolysed amino acids in aquatic biological samples

Open Access
|Mar 2019

Abstract

A sensitive and widely applicable High-Performance Liquid Chromatography (HPLC) with fluorescence detector (FLD) method for determination hydrolysed amino acids in fish and aquatic products was developed and validated through study. The method involved a microwave-assisted sample extraction and protein hydrolysed with the hydrochloric acid and propionic acid. The separation was performed by Shim-pack HR-ODS column with a gradient mobile phase. The total separation time was 25 min per run. The method showed satisfactory linearity, repeatability, accuracy, and limit of detection for 21 amino acids with the 1-100 mg/L calibration range based on 10 µL injection. The fish, fish feed, canned fish and seaweeds samples were analysed successfully by using this developed method.
Language: English
Page range: 17 - 23
Published on: Mar 15, 2019
Published by: Sri Lanka Association of Fisheries and Aquatic Resources
In partnership with: Paradigm Publishing Services

© 2019 G.D.T.M. Jayasinghe, B.K.K.K. Jinadasa, published by Sri Lanka Association of Fisheries and Aquatic Resources