Skip to main content
Have a personal or library account? Click to login
Biomechanical analysis of diamond-like carbon coated growth-guidance spinal system under cyclic axial compression Cover

Biomechanical analysis of diamond-like carbon coated growth-guidance spinal system under cyclic axial compression

Open Access
|Sep 2026

Figures & Tables

Figure 1

Set for four-segmental stabilization of scoliosis in children SOCORE GGS by NovaSpine: (a) diagram of installation of elements in the Th12-L6 spine segment with transverse connector (green colour); and (b) typical fixed pedicle screw with locking cap (orange colour) and sliding cap of pedicle screw with a gap for the movement of the rod (blue colour) with set (or cap) screw without coatings (NC) and with DLC coatings.

Figure 2

Schematic cross-sectional view of the coated sliding cap (sliding nut) showing the Ti-6Al-4V substrate, CrN interlayer (20 nm), and outer a-C-H(Si) DLC coating (1,000 nm). The layer thicknesses are not shown to scale.

Figure 3

Schematic diagram of: (a) measurement system for high-cycle compression tests; (b) placement of markers in the Th12–L6 vertebral bodies, based on which changes in the angular position of the vertebrae were determined in the sagittal and horizontal planes; and (c) measurement system using digital image correlation with a speckle pattern applied to the implant surfaces.

Figure 4

Characteristics: (a) a change in the height value (H); and (b) value of the stiffness coefficient (k) in subsequent cycles for the groups C, NC, and WC. *– Statistical significance (p < 0.05).

Figure 5

Graph of angular range of motion, in the sagittal plane, of the Th12–L6 vertebrae during axial spine compression for the groups C, NC, and WC: (a) scheme of the arrangement of markers (red points) and stabilizer elements – screws with a sliding nut (blue), standard non-sliding screws (orange), connector (green); (b) initial value of angular displacement after 100 cycles; and (c) final value of angular displacement after 100,000 cycles. *– Statistical significance (p < 0.05).

Figure 6

Mechanical parameters of sliding pedicle screw (L5 vertebra) in NC and WC. The value were obtained using the DIC method after 100 and 100,000 cycles: (a) engineering principal strain X (%); (b) exemplary distribution engineering principal strain X (%) of NC stabilization; (c) engineering principal strain Y (%); and (d) value of the total displacement in groups NC and WC.

DOI: https://doi.org/10.37190/abb/225557 | Journal eISSN: 2450-6303 | Journal ISSN: 1509-409X
Language: English
Page range: 1 - 11
Submitted on: May 20, 2026
Accepted on: Jul 7, 2026
Published on: Sep 21, 2026
Published by: Wroclaw University of Science and Technology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2026 Małgorzata Żak, Sylwia Szotek, Klaudia Szkoda-Poliszuk, Krzysztof Kołtowski, Piotr Menartowicz, Celina Pezowicz, published by Wroclaw University of Science and Technology