Observatory Magnetometer In-Situ Calibration
By: A Marusenkov, A Chambodut, J -J Schott and V Korepanov
Abstract
An experimental validation of the in-situ calibration procedure, which allows estimating parameters of observatory magnetometers (scale factors, sensor misalignment) without its operation interruption, is presented. In order to control the validity of the procedure, the records provided by two magnetometers calibrated independently in a coil system have been processed. The in-situ estimations of the parameters are in very good agreement with the values provided by the coil system calibration.
DOI: https://doi.org/10.2481/dsj.IAGA-17 | Journal eISSN: 1683-1470
Language: English
Published on: Jul 5, 2011
Published by: Ubiquity Press
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year
Keywords:
© 2011 A Marusenkov, A Chambodut, J -J Schott, V Korepanov, published by Ubiquity Press