Have a personal or library account? Click to login
Observatory Magnetometer In-Situ Calibration Cover

Observatory Magnetometer In-Situ Calibration

Open Access
|Jul 2011

Abstract

An experimental validation of the in-situ calibration procedure, which allows estimating parameters of observatory magnetometers (scale factors, sensor misalignment) without its operation interruption, is presented. In order to control the validity of the procedure, the records provided by two magnetometers calibrated independently in a coil system have been processed. The in-situ estimations of the parameters are in very good agreement with the values provided by the coil system calibration.
DOI: https://doi.org/10.2481/dsj.IAGA-17 | Journal eISSN: 1683-1470
Language: English
Published on: Jul 5, 2011
Published by: Ubiquity Press
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2011 A Marusenkov, A Chambodut, J -J Schott, V Korepanov, published by Ubiquity Press
This work is licensed under the Creative Commons Attribution 4.0 License.