Analysis of the Failure of Fixator used in Bone Surgery
By: B. Świeczko-Żurek, W. Serbiński and A. Szumlański
Open Access
|Oct 2008Abstract
The Gamma nail (fixator) made of stainless steel, which has broken in the body, was subjected to examination. The fixator was removed from the patient femur. The fracture occurred in the area of the nail hole after two months' stay in the organism. The research included chemical composition, hardness of the Gamma nail material and the observation and analysis of the type of fracture. The research showed the fatigue failure of the Gamma nail.
Language: English
Page range: 84 - 88
Published on: Oct 17, 2008
Published by: Gdansk University of Technology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year
Keywords:
Related subjects:
© 2008 B. Świeczko-Żurek, W. Serbiński, A. Szumlański, published by Gdansk University of Technology
This work is licensed under the Creative Commons License.