Does CyberKnife improve dose distribution versus IMRT and VMAT on a linear accelerator in low-risk prostate cancer? Cover

Does CyberKnife improve dose distribution versus IMRT and VMAT on a linear accelerator in low-risk prostate cancer?

Open Access
|Mar 2022

Abstract

Background

Hypofractionated stereotactic body radiation therapy (SBRT) for prostate cancer (PCa) can be delivered with the robot-assisted CyberKnife (CK) system or on a linear accelerator using dynamic intensity-modulated radiotherapy (IMRT) or volumetric arc radiotherapy (VMAT). This retrospective study was performed to determine whether CK offers better dose distribution than IMRT and/or VMAT.

Materials and methods

Treatment plans for three techniques were prepared using the same treatment parameters (36.35 Gy, 7.25 Gy/fr). We evaluated target coverage, conformity index (CI), homogeneity index (HI), gamma index (GI), and organs at risk (OAR) constraints.

Results

The mean planning target volume (PTV) dose for CK (39.58 Gy) was significantly greater than VMAT or IMRT (both 36.25 Gy). However, CK resulted in a wider dose range (31.48 to 45.89 Gy) vs. VMAT and IMRT (34.6–38.76 Gy). The mean dose to the rectum (V36Gy, mm3) was significantly lower (p < 0.001) in the CK plans (219.78 vs. 519.59 and 422.62, respectively). The mean bladder dose (V37Gy, mm3) was significantly greater for CK (3256 vs. 1090.75 for VMAT and 4.5 for IMRT (p < 0.001). CK yielded significantly better CI (1.07 vs. 1.17 and 1.25 for VMAT and IMRT, respectively; p < 0.01) and HI values (1.27 vs. 1.07 and 1.04; p < 0.01). GI values for the δd = 3mm, δ% = 3% criteria were 99.86 (VMAT), 99.07 (IMRT) and 99.99 (CK). For δd = 2mm, δ% = 2%, the corresponding values were 98.3, 93.35, and 97.12, respectively.

Conclusions

For most variables, CK was superior to both VMAT and IMRT. However, dynamic IMRT techniques, especially VMAT, do not differ significantly from CK plans and are therefore acceptable alternatives to CyberKnife.

DOI: https://doi.org/10.2478/raon-2022-0010 | Journal eISSN: 1581-3207 | Journal ISSN: 1318-2099
Language: English
Page range: 259 - 266
Submitted on: Oct 21, 2021
Accepted on: Jan 31, 2022
Published on: Mar 28, 2022
Published by: Association of Radiology and Oncology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 times per year

© 2022 Dorota Maria Borowicz, Agnieszka Skrobała, Marta Kruszyna-Mochalska, Julian Malicki, published by Association of Radiology and Oncology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.