A protocol for accurate radiochromic film dosimetry using Radiochromic.com

Open Access
|Aug 2021

Abstract

Background

Radiochromic films have many applications in radiology and radiation therapy. Generally, the dosimetry system for radiochromic film dosimetry is composed of radiochromic films, flatbed scanner, and film analysis software. The purpose of this work is to present the effectiveness of a protocol for accurate radiochromic film dosimetry using Radiochromic.com as software for film analysis.

Materials and methods

Procedures for image acquisition, lot calibration, and dose calculation are explained and analyzed. Radiochromic.com enables state-of-the-art models and corrections for radiochromic film dosimetry, such as the Multigaussian model for multichannel film dosimetry, and lateral, inter-scan, and re-calibration corrections of the response.

Results

The protocol presented here provides accurate dose results by mitigating the sources of uncertainty that affect radiochromic film dosimetry.

Conclusions

Appropriate procedures for film and scanner handling in combination with Radiochromic.com as software for film analysis make easy and accurate radiochromic film dosimetry feasible.

DOI: https://doi.org/10.2478/raon-2021-0034 | Journal eISSN: 1581-3207 | Journal ISSN: 1318-2099
Language: English
Page range: 369 - 378
Submitted on: Jun 24, 2021
|
Accepted on: Jul 1, 2021
|
Published on: Aug 10, 2021
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2021 Ignasi Méndez, Juan José Rovira-Escutia, Bozidar Casar, published by Association of Radiology and Oncology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.