Experimental validation of Monte Carlo based treatment planning system in bone density equivalent media Cover

Experimental validation of Monte Carlo based treatment planning system in bone density equivalent media

Open Access
|Sep 2020

Figures & Tables

Figure 1

Photo of the semi-anthropomorphic CIRS Thorax phantom with interchangeable rod inserts (left) and its CT image (right). Positions of 10 interchangeable rod inserts are marked with numbers from 1 to 10. Five measuring points are in the water equivalent part of the phantom (grey area), four points are in the lung density equivalent material (black area), and one point is in the bone density equivalent part of the phantom (white area).

Table 1

Irradiation set-ups for measurements in 6 MV photon beam used for experimental verification of the Monaco ver. 5.11 treatment planning systems (TPS) calculation algorithm in the semi-anthropomorphic CIRS Thorax phantom. Reference and measuring points (I1 to I10) are shown in the last two columns; subscripts 1 to 10 correspond to the labelling in Figure 1

Set-upIrradiation geometryField size [cm2]SSD/SADGantry angle [°]reference pointmeasuring points
1Single square fields10×10SSD0I5I1, I3, I5-10
210×10SAD0I5I, I, I135-10
34×4SAD0I5I1-9
410×10SAD90I3I2-10
5Rectangular field10× 15SAD300I1I1, I4, I6-8, I10
6Single asymmetric fields(6+8)×15SAD0I5I1-10
7(3+8)×15SAD90I5I, I15-10
8(4+10)×15SAD180I5I1-3, I5-10
9(3+7)×15SAD300I5I2-10
12×10SAD0
104 fields (box)12×10SAD180I5I2-5
12×8SAD90
12×8SAD270
4×4SAD30
113 fields16×4SAD90I5I2, I5-9
16×4SAD270
12Diamond-shaped field14×14SAD0I3I1, I3, I5-10
13Irregular L shaped field/SAD45I1I1-2, I4-6, I8-10
14MLC cylinder shaped field/SAD0I2I1,2, I5, I8,9, I10
16×4SAD90
153 non-coplanar fields16×4SAD270I5I1, I5-6, I8, I10
4×4aSAD30

SAD = source to axis distance; SSD = source to surface distance

a Couch angle = 270°

Figure 2

CT image of the CIRS Thorax phantom: water equivalent insert inside BDE part of the phantom (A); a BDE insert inside bone density equivalent (BDE) part of the phantom (B) and cross-section of small “water cylinders” of different dimensions delineated inside BDE part of the phantom to find limits for calculating geometry where cavity theory applies (top right).

Figure 3

Mean percentage dose differences δD¯mand δD¯wbetween calculated and measured doses in different parts of the CIRS Thorax phantom (water, lung, and bone density equivalent materials) for both calculation options built in the Monaco TPS: dose-to-medium Dm and dose-to-water Dw. Error bars represent corresponding combined uncertainties.

Table 2

Differences δDmand δDwbetween two different calculation options in the Monaco ver. 5.11. treatment planning systems (TPS) and measured data obtained in the bone density equivalent (BDE) part of the CIRS Thorax phantom, according to Eqs. [1] and [2]. Two phantom assemblies and three simple beam setups were considered for this part of the study

Irradiation geometry (field, gantry)Phantom assemblyδDm[%]δDw[%]
(6+8) x 15 cm2
Gantry = 0°
standarda- 2.92.9
non-standardb- 0.7- 0.2
(3+8) x 15 cm2
Gantry = 90°
standarda- 3.05.1
non-standardb- 0.7- 0.1
(4+10) x 15 cm2
Gantry = 180°
standarda- 5.75.4
non-standardb0.51.3

a BDE insert with the ionization chamber placed in the BDE part of the phantom

b Water equivalent insert with the ionization chamber placed in the BDE part of the phantom

a BDE insert with the ionization chamber placed in the BDE part of the phantom

b Water equivalent insert with the ionization chamber placed in the BDE part of the phantom

a BDE insert with the ionization chamber placed in the BDE part of the phantom

b Water equivalent insert with the ionization chamber placed in the BDE part of the phantom

Table 3

Mean differences, δD¯mand δD¯wbetween calculated and measured doses in the bone density equivalent (BDE) part of the CIRS Thorax phantom for Dm and Dw calculation approaches, respectively. The absorbed doses were calculated using the Monaco ver. 5.11 treatment planning systems (TPS) in the center of delineated “water cylinders” of volume V, in the BDE part of the phantom. Corresponding combined uncertainties are denoted as uc,m and uc,w for dose-to-medium and dose-to-water calculation options, respectively

V [cm3]δD¯m[%]uc,m[%]δD¯w[%]uc,w[%]
0- 3.92.14.41.9
0.035- 2.61.52.51.9
0.141- 1.41.31.81.7
0.279- 0.31.21.21.5
0.5730.31.40.41.3
Figure 4

Average differences δD¯mand δD¯wbetween calculated and measured doses in the bone density equivalent (BDE) part of the CIRS Thorax phantom, as a function of the volumes of the simulated “water cylinders” (see Figure 2 and Table 3). δD¯mand δD¯ware presented as individual values/points calculated using Eqs. [1] to, [4], and in the form of two analytical functions from Eqs. [9] and [10]. Error bars represent corresponding uncertainties within 95% confidence limits.

DOI: https://doi.org/10.2478/raon-2020-0051 | Journal eISSN: 1581-3207 | Journal ISSN: 1318-2099
Language: English
Page range: 495 - 504
Submitted on: Jun 10, 2020
Accepted on: Jul 9, 2020
Published on: Sep 16, 2020
Published by: Association of Radiology and Oncology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2020 Djeni Smilovic Radojcic, Bozidar Casar, David Rajlic, Manda Svabic Kolacio, Ignasi Mendez, Nevena Obajdin, Dea Dundara Debeljuh, Slaven Jurkovic, published by Association of Radiology and Oncology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.