Analysis of 177Lu-DOTATATE injection methods and their impact on the exposure of hands of nuclear medicine facility staff Cover

Analysis of 177Lu-DOTATATE injection methods and their impact on the exposure of hands of nuclear medicine facility staff

Open Access
|Sep 2026
Read online

Abstract

Purpose

The increasing use of peptide receptor radionuclide therapy (PRRT) with 177Lu-DOTATATE increases the need to assess occupational radiation exposure of nuclear medicine staff, particularly to the hands. This study aimed to evaluate and compare fingertip exposure associated with three administration methods: gravitational infusion, infusion pump, and manual injection.

Materials and Methods

Dosimetric measurements were performed in two nuclear medicine centres using LiF:Mg,Cu,P thermoluminescent detectors placed on the fingertips and at the base of the middle finger of both hands of staff involved in radiopharmaceutical preparation and administration. Measurements were collected for physicists, radiopharmacists, nurses, and physicians. A total of 126 measurements were collected in four measurement sessions. The personal dose equivalent Hp(0.07) was determined, normalised to administered activity (µSv/GBq), and analysed separately for preparation and administration procedures.

Results

Manual administration resulted in the highest fingertip exposure, with maximum normalized Hp(0.07) values of 26.95 µSv/GBq during preparation and 40.23 µSv/GBq during administration. Infusion pump administration showed intermediate exposure levels, while gravitational infusion yielded the lowest values, often close to the minimum detectable dose. Dominant hand fingertips, particularly the thumb and index finger, were most exposed. Estimated annual doses remained well below the skin dose limit.

Conclusions

The administration technique significantly influences occupational hand exposure during 177Lu- DOTATATE procedures. Manual administration leads to the highest doses, whereas gravitational and infusion pump methods substantially reduce staff exposure.

DOI: https://doi.org/10.2478/pjmpe-2026-0015 | Journal eISSN: 1898-0309 (formerly 1425-4689) | Journal ISSN: 1425-4689
Language: English
Page range: 138 - 145
Submitted on: Feb 3, 2026
Accepted on: Aug 6, 2026
Published online: Sep 30, 2026
Published in print: Sep 1, 2026
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services

© 2026 Łukasz Albiniak, Hanna Piwowarska-Bilska, Marta Rowińska, Bożena Birkenfeld, Małgorzata Wrzesień, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.