Feasibility of output quality assurance considering gantry angle using “Stealth Chamber” Cover

Feasibility of output quality assurance considering gantry angle using “Stealth Chamber”

Open Access
|Dec 2021

Abstract

Purpose: This study was conducted to demonstrate the feasibility of X-ray output constancy quality assurance (QA) of a linear accelerator for various gantry angles using the Stealth Chamber.

Methods: The X-ray output constancy of a Varian TrueBeam STx was evaluated under various gantry angles and a 10 × 10 cm2 field size using a Stealth Chamber. Specifically, 10X and 10X-flattening-filter-free beams with dose rates of 600 and 2400 monitor units (MU)/min, respectively, were used. The Stealth Chamber was attached to the gantry head, and irradiation was performed every 45° for gantry angles of 0-315°. To evaluate the variations in the output constancy with respect to the gantry angle, the acquired values were normalized to the value corresponding to a 0° gantry angle. The obtained results were utilized to determine the correction factors for all gantry angles. To verify the correction factors, additional measurements were performed for five days.

Results: The maximum variation in the output constancy measurement relative to the output constancy at a 0° gantry angle was found to be approximately 4.0% for both energy beams at a gantry angle of 180°. Furthermore, the measured values were dependent on the gantry angle. Upon applying the correction factor, the variation in the output constancy with respect to the gantry angle was less than 0.5%.

Conclusions: Output constancy QA using the Stealth Chamber for various gantry angles was found to be feasible with the application of a correction factor.

DOI: https://doi.org/10.2478/pjmpe-2021-0035 | Journal eISSN: 1898-0309 | Journal ISSN: 1425-4689
Language: English
Page range: 299 - 302
Published on: Dec 23, 2021
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2021 Hideharu Miura, Fumika Takeichi, Tsubasa Enosaki, Kiyoshi Yamada, Shuichi Ozawa, Yasushi Nagata, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.