First Evaluation of the PTN-104 Plethysmographic Sensor for Heart Rate Measurement Cover

First Evaluation of the PTN-104 Plethysmographic Sensor for Heart Rate Measurement

Open Access
|Sep 2021

Abstract

The purpose of this study was to examine the accuracy of HR measurements by the PTN-104 sensor in comparison to the fingertip pulse oximeter, which is a photoplethysmographic sensor (PPG). Twelve healthy participants underwent the same protocol during a single visit. Measurements were taken after each participant completed an initial rest period of 5 minutes and after 1-minute of exercising comprising of 30 squats. Each subject had the PTN-104 sensor attached to the index finger and a fingertip pulse oximeter to the opposite one. When examining the data in aggregate, there was a strong correlation between the PTN-104 sensor and PPG for HR (r = 0.988) with a mean bias of −2.55 bpm (95 % LoA +5.0, −10.1). The PTN-104 sensor satisfied validity criteria for HR monitors, however, showed a lower accuracy for measurements at rest, which is surprising. Due to the noticed limitations, this study should be repeated with a larger group of subjects and the PTN-104 sensor should be compared to the gold standard method for measuring HR, which is ECG.

Language: English
Page range: 117 - 122
Submitted on: Apr 29, 2021
|
Accepted on: Aug 26, 2021
|
Published on: Sep 17, 2021
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2021 Wiesław Szaj, Wiktoria Wojnarowska, Bogdan Pajdo, published by Slovak Academy of Sciences, Institute of Measurement Science
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.