Safety and Efficacy of Magnet Use to Temporarily Inhibit Inappropriate Subcutaneous Implantable Cardioverter Defibrillator Therapy in Emergency Situations: A Case Report
Abstract
Introduction: The subcutaneous implantable cardioverter defibrillator (S-ICD) represents a major advancement in ICD technology. Inappropriate shocks (IAS) occur in more than 3.1% of the population with S-ICD each year and are usually followed by admission to the emergency department (ED). In this setting, the disabling of IAS is mandatory during a pseudo-electrical storm (ES). This report describes the strategies that can be followed in order to temporarily inhibit IAS in critical care settings with the use of magnets.
Case presentation: An S-ICD was implanted more than 6 weeks prior to presentation in a 68-year-old man with hypertrophic cardiomyopathy. In the ED, the patient experienced 3 IAS in the last hour. A Medtronic magnet was applied to stop IAS, as the specific programmer was not available. The maneuver interrupted the IAS. In order to verify the universal magnetic response of the S-ICD, six different magnets and one smartphone with MagSafe technology were tested. All magnet models suspended arrhythmia detection and IAS, while the smartphone did not cause magnet interferences.
Conclusions: This report demonstrates the safety and efficacy of all clinical magnet models in inhibiting IAS. In case of pseudo-ES, any type of magnet allows ED providers to easily and rapidly disable the functionality of the devices when appropriate.
© 2022 Maurizio Santomauro, Carla Riganti, Mario Alberto Santomauro, Aniello Viggiano, Gaetano Castellano, Gianluigi Iovino, Antonio Rapacciuolo, Francesco Fiore, Francesco Cacciatore, Giovanni Esposito, published by Asociatia Transilvana de Terapie Transvasculara si Transplant KARDIOMED
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.