Skip to main content
Have a personal or library account? Click to login
Effect of urine adulterants on commercial drug abuse screening test strip results Cover
Read online

Figures & tables

Table 1

Adulterants considered for testing and the quantities used for those that passed urine appearance test

Household adulterantFormQuantity used
Acids
        WC Sanitar (9 % HCl)0.3 mL
        Vinegar (white distilled vinegar; 9 % acetic acid)liquid0.2 mL
        Lemon juice0.4 mL
        Citric acidsolid0.1 g
Alkalis
        Cevtok (30 % NaOH)liquid0.1 mL
Oxidizing agentsliquid
        Varikina (bleach; 5 % NaOCl)0.3 mL
Surfactants
        Asepsol (5 % quaternary ammonium compounds, such as benzalkonium chloride)0.2 mL
        Visine® eye drops (active substance: tetrahydrozoline-hydrochloride 0.05 % w/v; non-medical ingredients: benzalkonium-chloride, sodium chloride, boric acid, sodium borate, disodium EDTA)liquid0.4 mL
        Liquid soap*
        Household cleaning products Frosh, Aro and Skala*
Miscellaneous substances
        Primosept GA (8 % glutaraldehyde)liquid0.4 mL
        Zn salts (tablets, supplements)*solid
        Nitrite and nitrate salts (meat product additives)*solid

[i] * household chemicals that changed urine appearance and were not tested further for adulteration

Table 2

Urine drug (metabolite) concentrations and UDST results before adulteration with household chemicals

Drug (metabolite)Urine concentration (mg/L)UDST result
MDMA0.83+
Amphetamine2.28+
Cocaine (benzoylecgonine)5.69+
Diazepam0.077+
Alprazolam0.065+
Tetrahydrocannabinol (carboxy-tetrahydrocannabinol)112.2+
Heroine (6-acetylmorphine)0.23+

[i] “+” – test positive result

Table 3

UDST results obtained after adulteration of drug-positive urine samples with household chemicals

OpiatesCannabinoidsAmphetamineMDMACocaineAlprazolamDiazepam
WC SanitarInvalid-InvalidInvalid+InvalidInvalid
Vinegar----+--
Lemon juice+---+--
Citric acid+-Invalid++InvalidInvalid
CevtokInvalidInvalidInvalidInvalidInvalidInvalidInvalid
Varikina---++++
Asepsol+-InvalidInvalidInvalidInvalidInvalid
Primosept GA++InvalidInvalidInvalidInvalidInvalid
Visine+++++++

[i] “+” – test remained positive; “-” – test became negative; “Invalid” – omitting of control line or both lines on test strip

Table 4

Effects of urine adulteration on parameters investiated by URIT 11G urine reagent strips and pH test strips

URIT 11G urine reagent stripspH strips test
Ascorbic acid (mmol/L)Blood (cells/μL)Specific gravityGlucose (mmol/L)Protein g/LBilirubin (μmol/L)UrobilinogenNitrite (μmol/L)Ketone (mmol/L)Leukocytes (cells/μL)pHpH
Test strip sensitivity0.6-0.855-15/2.2-2.80.1-0.38.6-1717-3318-330.5-115-0//
Test strip range0-5.60-2001.005-1.0300-550-30-100Normal-131+/-0-80-5005-90-14
Control urine*001.0200-2.800-8.6Normal---65
Control urine +
added adulterant**
WC Sanitar2.8Invalid1.030000Normal---51
Vinegar2.801.0300-2.800Normal---54
Lemon juice2.8251.0305.50.158.6Normal---53
Citric acid2.8101.0300-2.800Normal---51
CevtokInvalid0Invalid0Invalid8.6Invalid--70>914
Varikina0801.0150-2.800Normal+--65
Asepsol001.0100Invalid0Normal---65
Primo sept GA001.03000.150Normal->8.0-55
Visine001.025000Normal---55

[i] "-" no presence detected; Invalid - malfunction of test; Normal - within the range of normal physiological parameters; bold results indicate invalid test, pathophysiological conditions or urine manipulation; *similar results were obtained in all tested drag positive urine samples before adulteration; "similar results were obtained in all tested drag positive urine samples after adulteration

DOI: https://doi.org/10.2478/aiht-2020-71-3315 | Journal eISSN: 1848-6312 (formerly 0004-1254) | Journal ISSN: 0004-1254
Language: English
Page range: 87 - 93
Submitted on: Jul 1, 2019
Accepted on: Mar 1, 2020
Published online: Apr 9, 2020
Published in print: Mar 1, 2020
Published by: Institute for Medical Research and Occupational Health
In partnership with: Paradigm Publishing Services

© 2020 Ivana Rajšić, Dragana Javorac, Simona Tatović, Aleksandra Repić, Danijela Đukić-Ćosić, Snežana Đorđević, Vera Lukić, Zorica Bulat, published by Institute for Medical Research and Occupational Health
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.