Skip to main content
Have a personal or library account? Click to login
Physiological bending sensor based on tilt angle loss measurement using dual optical fibre Cover

Physiological bending sensor based on tilt angle loss measurement using dual optical fibre

By: ,   and    
Open Access
|Feb 2020
Read online

Abstract

This paper presents the development of an extrinsic optical fibre sensor for continuous measurement of the human spine bending movement based on intensity modulation technique. Using the investigated sensor configuration, the bending angle was measurable in both flexion and extension direction with a maximum range of motion of 18° and -10°, respectively. From the output drift assessment of the sensor, bending accuracy of up to 0.5° was achievable, thus making it suitable for clinical environment application. A divided-beam referencing technique was also implemented in the sensor configuration to compensate for the input power fluctuation and temperature variation.

Language: English
Page range: 1 - 4
Published online: Feb 15, 2020
Published in print: Jan 1, 2014
Published by: International Journal on Smart Sensing and Intelligent Systems
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2020 M. A. Zawawi, S. O’Keeffe, E. Lewis, published by International Journal on Smart Sensing and Intelligent Systems
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.