Have a personal or library account? Click to login

Calibration Method for the Wavelength and Uniformity of Pixel Response in Photodetector Arrays

Open Access
|Sep 2018

Figures & Tables

Fig. 1

Camera—FL3-U3-13Y3M-C.
Camera—FL3-U3-13Y3M-C.

Fig. 2

Experiment of calibration.
Experiment of calibration.

Fig. 3

Defective pixels in the array.
Defective pixels in the array.

Fig. 4

Uniformity.
Uniformity.

Fig. 5

Output monochromator—experimental result.
Output monochromator—experimental result.

Fig. 6

Sensor response camera.
Sensor response camera.

Mean response and margin_

Wavelength (nm)P1 (%)P2 (%)P3 (%)P4 (%)P5 (%)Margin (%)
38053.3453.9553.5653.7853.700.62
40063.3361.8665.2365.4463.523.58
42071.1570.6472.8270.7469.753.07
44077.5075.9275.3376.0574.952.55
46079.7082.5182.8381.3479.453.39
48084.3584.7386.2583.2383.533.02
50087.9889.8089.7889.4588.781.82
52093.6192.6494.4092.2793.872.13
54095.1395.9096.9197.4594.782.67
56099.8799.3598.2998.0498.071.83
58098.0298.7998.84100.00100.001.98
60099.2698.2399.8098.7198.071.72
62099.7399.6798.3498.6898.501.39
64094.8495.3895.8095.0597.372.53
66089.9890.0989.9390.1391.011.09
68086.7089.2687.2286.2789.002.98
70085.7383.6682.4683.0183.673.27
72078.4777.2278.2378.7276.971.74
74069.5370.4770.3270.9571.121.60
76063.6164.3266.5063.6766.723.11
78060.2659.4059.7858.4359.261.83
Language: English
Page range: 1 - 8
Published on: Sep 17, 2018
Published by: Professor Subhas Chandra Mukhopadhyay
In partnership with: Paradigm Publishing Services
Publication frequency: 1 times per year

© 2018 J.-S. Botero-Valencia, J. Valencia-Aguirre, D.A. Marquez-Viloria, J.-F. Vargas-Bonilla, A. Gallardo-del Angel, published by Professor Subhas Chandra Mukhopadhyay