Have a personal or library account? Click to login
Calibration Method for the Wavelength and Uniformity of Pixel Response in Photodetector Arrays Cover

Calibration Method for the Wavelength and Uniformity of Pixel Response in Photodetector Arrays

Open Access
|Sep 2018

Figures & Tables

Fig. 1

Camera—FL3-U3-13Y3M-C.
Camera—FL3-U3-13Y3M-C.

Fig. 2

Experiment of calibration.
Experiment of calibration.

Fig. 3

Defective pixels in the array.
Defective pixels in the array.

Fig. 4

Uniformity.
Uniformity.

Fig. 5

Output monochromator—experimental result.
Output monochromator—experimental result.

Fig. 6

Sensor response camera.
Sensor response camera.

Mean response and margin_

Wavelength (nm)P1 (%)P2 (%)P3 (%)P4 (%)P5 (%)Margin (%)
38053.3453.9553.5653.7853.700.62
40063.3361.8665.2365.4463.523.58
42071.1570.6472.8270.7469.753.07
44077.5075.9275.3376.0574.952.55
46079.7082.5182.8381.3479.453.39
48084.3584.7386.2583.2383.533.02
50087.9889.8089.7889.4588.781.82
52093.6192.6494.4092.2793.872.13
54095.1395.9096.9197.4594.782.67
56099.8799.3598.2998.0498.071.83
58098.0298.7998.84100.00100.001.98
60099.2698.2399.8098.7198.071.72
62099.7399.6798.3498.6898.501.39
64094.8495.3895.8095.0597.372.53
66089.9890.0989.9390.1391.011.09
68086.7089.2687.2286.2789.002.98
70085.7383.6682.4683.0183.673.27
72078.4777.2278.2378.7276.971.74
74069.5370.4770.3270.9571.121.60
76063.6164.3266.5063.6766.723.11
78060.2659.4059.7858.4359.261.83
Language: English
Page range: 1 - 8
Published on: Sep 17, 2018
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2018 J.-S. Botero-Valencia, J. Valencia-Aguirre, D.A. Marquez-Viloria, J.-F. Vargas-Bonilla, A. Gallardo-del Angel, published by Professor Subhas Chandra Mukhopadhyay