Have a personal or library account? Click to login
Acoustic radiation force impulse imaging of kidneys – a phantom study Cover

Acoustic radiation force impulse imaging of kidneys – a phantom study

Open Access
|Feb 2017

Abstract

Aim of the study:

Since there have been only few works reporting the diagnosis of kidneys using Acoustic Radiation Force Impulse technique and those works do not provide consistent results of shear wave velocity measurements in renal tissue, we have decided to use kidney phantoms with known properties to examine the reliability of the method itself in a controlled setup similar to kidneys examination.

Materials and methods:

Four gelatin-based phantoms imitating different clinical situations were manufactured – two with thick and two with thin renal cortex, each type at a depth similar to a normal-weight or overweight patient. For each phantom, a series of interest points was chosen and for each point 20 Shear Wave Velocity measurements were taken using the build-in Virtual Touch Tissue Quantification™tool in a Siemens Acuson S2000 ultrasound scanner equipped with a 6C1 HD Transducer (Siemens Mountainview, USA).

Results:

Mean Shear Wave Velocity values obtained for all the examined points ranged from 2.445 to 3.941 m/s, with standard deviation exceeding 0.1 in only one case out of 29 points, but differing significantly between all points.

Conclusions:

The obtained results indicate that the method is highly reliable as long as the measurement volume contains a uniform tissue region. If the measurement window covers a region with different properties even partially, the obtained results are affected. The variance of measured values on the other hand is not affected by the said non-uniformity of material under examination. Furthermore, the variance of measured values does not show a clear dependency on the depth at which the shear wave velocities are measured.

DOI: https://doi.org/10.15557/jou.2016.0033 | Journal eISSN: 2451-070X | Journal ISSN: 2084-8404
Language: English
Page range: 5 - 8
Submitted on: Nov 9, 2015
Accepted on: Mar 14, 2016
Published on: Feb 24, 2017
Published by: MEDICAL COMMUNICATIONS Sp. z o.o.
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2017 Szymon Cygan, Magdalena Januszewicz, published by MEDICAL COMMUNICATIONS Sp. z o.o.
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.