Have a personal or library account? Click to login
Optimal planning strategy among various arc arrangements for prostate stereotactic body radiotherapy with volumetric modulated arc therapy technique Cover

Optimal planning strategy among various arc arrangements for prostate stereotactic body radiotherapy with volumetric modulated arc therapy technique

Open Access
|Jan 2017

Abstract

Background

The aim of this study was to determine the optimal strategy among various arc arrangements in prostate plans of stereotactic body radiotherapy with volumetric modulated arc therapy (SBRT-VMAT).

Patients and methods

To investigate how arc arrangements affect dosimetric and biological metrics, SBRT-VMAT plans for eighteen patients were generated with arrangements of single-full arc (1FA), single-partial arc (1PA), double-full arc (2FA), and double-partial arc (2PA). All plans were calculated by the Acuros XB calculation algorithm. Dosimetric and radiobiological metrics for target volumes and organs at risk (OARs) were evaluated from dosevolume histograms.

Results

All plans were highly conformal (CI<1.05, CN=0.91) and homogeneous (HI=0.09-0.12) for target volumes. For OARs, there was no difference in the bladder dose, while there was a significant difference in the rectum and both femoral head doses. Plans using 1PA and 2PA showed a strong reduction to the mean rectum dose compared to plans using 1FA and 2FA. Contrastively, the D2% and mean dose in both femoral heads were always lower in plans using 1FA and 2FA. The average tumor control probability and normal tissue complication probability were comparable in plans using all arc arrangements.

Conclusions

The use of 1PA had a more effective delivery time and produced equivalent target coverage with better rectal sparing, although all plans using four arc arrangements showed generally similar for dosimetric and biological metrics. However, the D2% and mean dose in femoral heads increased slightly and remained within the tolerance. Therefore, this study suggests that the use of 1PA is an attractive choice for delivering prostate SBRT-VMAT.

DOI: https://doi.org/10.1515/raon-2017-0005 | Journal eISSN: 1581-3207 | Journal ISSN: 1318-2099
Language: English
Page range: 112 - 120
Submitted on: Jul 19, 2016
Accepted on: Oct 25, 2016
Published on: Jan 15, 2017
Published by: Association of Radiology and Oncology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2017 Sang Won Kang, Jin Beom Chung, Jae Sung Kim, In Ah Kim, Keun Yong Eom, Changhoon Song, Jeong Woo Lee, Jin Young Kim, Tae Suk Suh, published by Association of Radiology and Oncology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.