Have a personal or library account? Click to login
The Effects of Fluid Viscosity on the Orifice Rotameter Cover

The Effects of Fluid Viscosity on the Orifice Rotameter

Open Access
|May 2016

Abstract

Due to the viscous shear stress, there is an obvious error between the real flow rate and the rotameter indication for measuring viscous fluid medium. At 50 cSt the maximum error of DN40 orifice rotameter is up to 35 %. The fluid viscosity effects on the orifice rotameter are investigated using experimental and theoretical models. Wall jet and concentric annulus laminar theories were adapted to study the influence of viscosity. And a new formula is obtained for calculating the flow rate of viscous fluid. The experimental data were analyzed and compared with the calculated results. At high viscosity the maximum theoretical results error is 6.3 %, indicating that the proposed measurement model has very good applicability.

Language: English
Page range: 87 - 95
Submitted on: Nov 11, 2015
Accepted on: Apr 5, 2016
Published on: May 6, 2016
Published by: Slovak Academy of Sciences, Institute of Measurement Science
In partnership with: Paradigm Publishing Services
Publication frequency: Volume open

© 2016 Wei Jiang, Tao Zhang, Ying Xu, Huaxiang Wang, Xiaoli Guo, Jing Lei, Peiyong Sang, published by Slovak Academy of Sciences, Institute of Measurement Science