Thorax mapping for localized lung impedance change using focused impedance measurement (FIM): A pilot study Cover

Thorax mapping for localized lung impedance change using focused impedance measurement (FIM): A pilot study

Open Access
|Dec 2013

Abstract

Focused impedance measurement (FIM) is a technique where impedance can be measured with the optimum level of localization without much increase in complexity of measuring instrument. The electrodes are applied on the skin surface while the organs inside also contribute to the measurement, as the body is a volume conductor. In a healthy and disease free lung region, the air enters at breathe-in, increases the impedance of the lung, and impedance reduces during breathe-out. In contrast, for a diseased lung, where part of the lungs is filled with water or some fluid, air will not enter into this zone reducing impedance change between inspiration and expiration. With this idea, the current work had been executed to have general view of localized impedance change throughout thorax using 6-electrode FIM. This generated a matrix mapping from both the front and from the back of the thorax, which showed how impedance change due to ventilation varies from frontal plane to back plane of human bodies.

DOI: https://doi.org/10.5617/jeb.778 | Journal eISSN: 1891-5469
Language: English
Page range: 57 - 61
Submitted on: Dec 4, 2013
Published on: Dec 28, 2013
Published by: University of Oslo
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2013 Humayra Ferdous, Tanvir Noor Baig, K. Siddique-e Rabbani, published by University of Oslo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.