Have a personal or library account? Click to login
Impedimetric characterization of human blood using three-electrode based ECIS devices Cover

Impedimetric characterization of human blood using three-electrode based ECIS devices

Open Access
|Jul 2012

Figures & Tables

Fig. 1

Schematic representation of three-electrode based ECIS devices.
Schematic representation of three-electrode based ECIS devices.

Fig. 2

Microphotograph of different three-electrode devices. Inset is the enlarged view of Design 4.
Microphotograph of different three-electrode devices. Inset is the enlarged view of Design 4.

Fig. 3

Bode Plot of PBS with its equivalent circuit.
Bode Plot of PBS with its equivalent circuit.

Fig. 4

Bode Plot of Blood with its equivalent circuit.
Bode Plot of Blood with its equivalent circuit.

Fig.5

Bode Plot of serum and plasma with its equivalent circuit.
Bode Plot of serum and plasma with its equivalent circuit.

Fig. 6

Bode Plot of RBCs with its equivalent circuit.
Bode Plot of RBCs with its equivalent circuit.

Fig. 7

Bode Plot of blood, serum, plasma, and RBCs for (1) Deign 1; (2) Design 2; (3) Design 3; (4) Design 4.
Bode Plot of blood, serum, plasma, and RBCs for (1) Deign 1; (2) Design 2; (3) Design 3; (4) Design 4.

Dimensions of different design of three electrode devices_

DevicesElectrode dimensions (μm2)Distance between electrodes ( μm)
Design 1WE- 50×50, RE- 500×500, CE- 500×500100
Design 2WE- 100×100, RE-1000×1000, CE- 1000×1000100
Design 3WE- 150×150, RE-1500×1500, CE- 1500×1500100
Design 4WE- 200×200, RE-2000×2000, CE- 2000×2000100

Values of electric components of RBC for different designs with relative standard error (e %)_

ECsDesign 1Design 2Design 3Design 4
Datae %Datae %Datae %Datae %
RS (Ω)9915.767744.246045.814455.9
CS (Fsn-1) ×10-920.26.5630.325.233.16.5649.56.94
RI (Ω)22252.9816512.413712.989983.21
QM (Fsn-1) ×10-312899.2915837.820469.3268110.6
n0.800.590.61±0.450.600.410.610.46
value χ23.7×10-32.62×10-33.8×10-34.79×10-3

Values of electric components of plasma and serum for different designs with relative standard error (e %)_

ECsDesign 1Design 2Design 3Design 4
Datae %Datae %Datae %Datae %
Plasma
RS (Ω)7610.85550.83950.782940.75
CS(Fsn-1) ×10-982.23112.13152.07202
n0.910.020.910.020.910.020.910.02
RCT (kΩ)12246.58746.16165.84625.6
χ2 value8.45×10-47.74×10-47.28×10-46.79×10-4
Serum
RS (Ω)8330.735960.683950.652800.62
CS(Fsn-1)×10-97.331.889.901.7714.81.720.61.61
n0.900.020.900.020.900.020.900.02
RCT (kΩ)8823.466263.224153.082922.90
χ2 value5.97×10-45.28×10-44.83×10-44.36×10-4

Values of electric components of blood for different designs with relative standard error (e %)

ECsDesign 1Design 2Design 3Design 4
Datae %Datae %Datae %Datae %
RS (Ω)13531.8311121.85906.51.79654.81.74
CS (Fsn-1) ×10-946.124.9855.135.0166.434.8190.194.64
(RΩI)15883.0113062.9910602.817672.67
QM (Fsn-1) ×10-95.053.166.143.167.52310.362.87
n0.920.030.920.030.920.030.920.03
χ2 value1.06×10-31.06×10-39.66×10-48.96×10-4
DOI: https://doi.org/10.5617/jeb.238 | Journal eISSN: 1891-5469
Language: English
Page range: 12 - 19
Submitted on: Jan 16, 2012
Published on: Jul 28, 2012
Published by: University of Oslo
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2012 Rangadhar Pradhan, Analava Mitra, Soumen Das, published by University of Oslo
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.