Abstract

Electrical impedance measurements of a suspension have to take into account the double layer impedance that results from a very thin charged layer formed at the electrode-electrolyte interface. A dedicated measuring cell that enables variation of the distance between the electrodes was developed to investigate the electrical properties of suspensions using two-electrode impedance measurements. By varying the distance between the electrodes it is possible to separate the double layer and the suspension impedance from the measured data. Electrical ‘lumped’ models have been developed from measured and extracted impedances. The error of non-inclusion of the double layer impedance has been analyzed. The error depends on the frequency of the measurement as well as the distance between the electrodes.

DOI: https://doi.org/10.5617/jeb.215 | Journal eISSN: 1891-5469
Language: English
Page range: 42 - 50
Submitted on: Nov 23, 2011
Published on: Oct 21, 2012
Published by: University of Oslo
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2012 Dejan Križaj, Borut Pečar, published by University of Oslo