Skip to main content
Have a personal or library account? Click to login
Hypoxic incident in the operating theatre due to a central oxygen pipeline problem Cover

Hypoxic incident in the operating theatre due to a central oxygen pipeline problem

By:   
Open Access
|Jul 2015

Abstract

Central oxygen pipeline failure is not a common problem. In our theatre complex compressed air got into the oxygen line while five surgeries were going on under general anaesthesia, causing hypoxia. The problem was identified in time and corrective measures were taken before any harm was done to the patients. The way in which this freak accident happened is described.

Language: English
Page range: 77 - 79
Published on: Jul 22, 2015
Published by: College of Anaesthesiologists of Sri Lanka
In partnership with: Paradigm Publishing Services

© 2015 AA Abraham, published by College of Anaesthesiologists of Sri Lanka
This work is licensed under the Creative Commons License.