Have a personal or library account? Click to login
Biomechanical assessment of lumbar stability: finite element analysis of TLIF with a novel combination of coflex and pedicle screws Cover

Biomechanical assessment of lumbar stability: finite element analysis of TLIF with a novel combination of coflex and pedicle screws

By: S. Meganathan and  M.S. Alphin  
Open Access
|Mar 2024

Abstract

Purpose: Finite element analysis is frequently used for lumbar spine biomechanical analysis. The primary scope of this work is to illustrate, using finite element analysis, how the biomechanical behavior of the transforaminal lumbar interbody fusion (TLIF), along with a novel combination of the interspinous process device (IPD) and pedicle screws, improves lumbar spine stability.

Methods: In this study, unilateral pedicle screw fixation (UPSF) and bilateral pedicle screw fixation (BPSF) were used. Four FE models were developed using ANSYS software, as follows: (1) Intact model; (2) TLIF with “U”-shaped Coflex-F IPD (UCF); (3) TLIF with Coflex-F and UPSF (UCF + UPSF); (4) TLIF with Coflex-F and BPSF (UCF + BPSF). The intact model was subjected to four pure moments (10 Nm), and the results were validated with previous literature data. The intact model results correlated well with the literature data, and the model was validated. Three surgical models were subjected to 7.5 Nm four pure moments, flexion (FL), extension (ET), lateral bending (LB), and axial rotation (AR) and a 280N follower load.

Results: The surgical model results were compared with the intact model. The comprehensive analysis results show the UCF + BPSF surgical model gave a good advantage on range of motion, cage stress, Coflex-F stress and endplate stress compared among the two models.

Conclusion: This study proposes that the UCF + BPSF system helps to reduce the stress on the implant and adjacent endplates and gives very good stability to the lumbar spine under the various static loading conditions.

DOI: https://doi.org/10.37190/abb-02380-2024-04 | Journal eISSN: 2450-6303 | Journal ISSN: 1509-409X
Language: English
Page range: 133 - 143
Submitted on: Jan 11, 2024
Accepted on: Mar 20, 2024
Published on: Mar 20, 2024
Published by: Wroclaw University of Science and Technology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2024 S. Meganathan, M.S. Alphin, published by Wroclaw University of Science and Technology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.