Calculations of carrier gas effects in non-dispersive infrared analyzers I. Theory
Abstract
A theory and method of calculation is presented for determining changes in instrument response and hence apparent gas concentration which arise in non-dispersive infrared gas analyzers when gas mixtures with varying carrier gases are analyzed. The theory is based on variations in pressure-broadening in the sample cell of the analyzer and initial comparisons with experimental values for CO2-in-air and CO2-in-N2 indicate good agreement between theory and experiment.
DOI: https://doi.org/10.3402/tellusa.v34i4.10824 | Journal eISSN: 3035-9554
Language: English
Page range: 376 - 384
Submitted on: Sep 21, 1981
Accepted on: Dec 23, 1981
Published on: Jan 1, 1982
Published by: Stockholm University Press
In partnership with: Paradigm Publishing Services
© 1982 David W. T. Griffith, published by Stockholm University Press
This work is licensed under the Creative Commons Attribution 4.0 License.