Errors in atmospheric CO2 concentration measurements arising from the use of reference gas mixtures different in composition to the sample air Cover

Errors in atmospheric CO2 concentration measurements arising from the use of reference gas mixtures different in composition to the sample air

By:  and    
Open Access
|Jan 1975

Abstract

Comparisons have been made between the URAS 2, UNOR 2 and Grubb Parson's SB2 non-dispersive infra-red gas analysers, involving the measurement of CO2 concentration in gas mixtures containing N2 and O2 in varying proportions. The results imply that the CO2 concentration in air when using CO2/N2 mixtures as references, is under-estimated by 3.7 and 2.3 ppmv for the URAS and Grubb Parson's instruments respectively and over-estimated by 4.5 ppmv by the UNOR. The measurement of CO2 concentration in air is not affected if CO2/air mixtures are used as references. The indicated CO2 concentration of air relative to a CO2/N2 reference mixture is found to depend on ambient pressure for both the UNOR and URAS analysers. Thus aircraft or high altitude in situ measurements need an additional correction, to be comparable with measurements made relative to the same CO2/N2 reference mixtures at a standard (near sea-level) pressure of 1 000 mb. The results are similar to those obtained by Bischof (1973) and have important implications on the interpretation of carbon dioxide concentration measurements made at monitoring stations.

Language: English
Page range: 62 - 66
Submitted on: Dec 12, 1973
Accepted on: Apr 24, 1974
Published on: Jan 1, 1975
Published by: Stockholm University Press
In partnership with: Paradigm Publishing Services

© 1975 G. I. Pearman, J. R. Garratt, published by Stockholm University Press
This work is licensed under the Creative Commons Attribution 4.0 License.