Skip to main content
Have a personal or library account? Click to login
Analysis of the measurement uncertainty during calibration of radioherapy and radiodiagnostics dosemeters at the Polish Secondary Standard Dosimetry Laboratory Cover

Analysis of the measurement uncertainty during calibration of radioherapy and radiodiagnostics dosemeters at the Polish Secondary Standard Dosimetry Laboratory

Open Access
|Apr 2012

Abstract

The aim of the study was to work out a method of establishing the measurement uncertainty and determination of uncertainty values for calibration of dosemeters at the SSDL. For the determination of type A and B uncertainties, a method according to iternational ISO standard was used. The calibration procedures used at the SSDL, consistent with the reports of the International Atomic Energy Agency, were taken into account. The presented results apply to all dosimeters calibrated at the SSDL. The detailed analysis of every step of the calibration procedures and the results obtained show that the measurement uncertinty is within the limits accepted by the Polish Centre for Accreditation.

DOI: https://doi.org/10.2478/v10013-011-0018-z | Journal eISSN: 1898-0309 (formerly 1425-4689) | Journal ISSN: 1425-4689
Language: English, Polish
Page range: 173 - 188
Published on: Apr 26, 2012
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services

© 2012 Piotr Ulkowski, Wojciech Bulski, Barbara Gwiazdowska, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons License.