Have a personal or library account? Click to login

Skin dosimetry of patients during interventional cardiology procedures in the Czech Republic

By:
Open Access
|Apr 2009

Abstract

The aim of the study is to determine distribution of air kerma-area product, fluoro time and number of frames values for the two most frequent procedures in the interventional cardiology, to reconstruct skin dose distributions for some patients undergoing coronarography and percutaneous transluminal coronary angioplasty procedures. Patient dose data were obtained from X-ray unit dose monitoring software report from one hospital and the reconstructions were performed in MATLAB. Dependence of maximum skin dose on air kerma-area product, fluoro time and number of frames was determined to assess trigger levels of these quantities, which can indicate possible exceeding of the 2 Gy skin dose threshold.

DOI: https://doi.org/10.2478/v10013-008-0018-9 | Journal eISSN: 1898-0309 | Journal ISSN: 1425-4689
Language: English
Page range: 217 - 225
Published on: Apr 14, 2009
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2009 Lucie Sukupova, Leos Novak, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons License.

Volume 14 (2008): Issue 4 (December 2008)