Evaluation of the potential in radiation dose reduction for full-field digital mammography Cover

Evaluation of the potential in radiation dose reduction for full-field digital mammography

By: Kay-Uwe Kasch and  Belal Moftah  
Open Access
|Apr 2009

Abstract

This study evaluates the image quality for different radiation doses in full-field digital mammography (FFDM). The potential of dose reductions is evaluated for both, the transition from screen-film mammography (SFM) to FFDM as well as within FFDM due to the optimization of exposure parameters.

Exposures of a 4.5 cm breast phantom rendering different contrasts as well as bar patterns were made using a FFDM system (GE Senographe 2000D). For different kVp and mAs settings as well as different target/filter combinations chosen for the above exposures, average glandular dose (AGD), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and modulation transfer function (MTF) were determined. To benchmark the results, relative change of AGD was evaluated against SNR, CNR and MTF. Eventually, the results were normalized to AGD's rendered by settings typically used in today's clinical routine.

For standard settings (automatic mode), both FFDM and SFM deliver approximately the same AGD of about 2.2 mGy. From that, AGD reduction can be substantial in FFDM if only SNR and high contrast CNR are considered. In this case, reduction of up to 40% can be achieved in a wide kVp range if switching from the standard target/filter combination Mo/Rh to Rh/Rh. However, if low contrast CNR is to remain unchanged, dose reduction is practically impossible. The change of peak voltage and target/filter material had no influence on MTF.

Assuming current CNR requirements as standards, significant dose reduction in FFDM cannot be achieved. Only by compromising low contrast CNR levels AGD of up to 40% can be saved at current standards of SNR and high contrast CNR.

DOI: https://doi.org/10.2478/v10013-008-0008-y | Journal eISSN: 1898-0309 | Journal ISSN: 1425-4689
Language: English
Page range: 87 - 98
Published on: Apr 14, 2009
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2009 Kay-Uwe Kasch, Belal Moftah, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons License.

Volume 14 (2008): Issue 2 (June 2008)