Have a personal or library account? Click to login
Ionisation Chambers Containing Boron as Neutron Detectors in Mixed Radiation Fields Cover

Ionisation Chambers Containing Boron as Neutron Detectors in Mixed Radiation Fields

Open Access
|Dec 2008

Abstract

The paper presents a newly designed ionisation chambers containing boron, operated in the initial recombination regime. The chambers were either filled with BF3 or the chamber electrodes were covered with B4C. The chambers can be placed in paraffin moderators. The sensitivity of the chambers was investigated depending on gas pressure, moderator thickness and polarizing voltage. The results showed that it was possible to obtain nearly the same sensitivity of the chamber to H* (10) for photons and neutrons in restricted energy range, however further investigations are needed to make an optimum design. The examples of applications for dosimetric measurements in mixed radiation fields near medical linear accelerator and in the vicinity of high-energy proton accelerator are presented.

DOI: https://doi.org/10.2478/v10013-007-0007-4 | Journal eISSN: 1898-0309 | Journal ISSN: 1425-4689
Language: English
Page range: 79 - 91
Published on: Dec 30, 2008
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2008 Michał Gryziński, Piotr Tulik, Mieczysław Zielczyński, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons License.

Volume 13 (2007): Issue 2 (June 2007)