Have a personal or library account? Click to login
Performance tests using the Lexsyg luminescence reader Cover

Performance tests using the Lexsyg luminescence reader

Open Access
|Oct 2014

Abstract

This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the type Standard, hence designed for routine determination of palaeodoses using quartz or K-feldspars. The tests include measurements of the stimulation powers, preheat temperatures, OSL- and TL-curves, as well as dose recovery tests and calibration measurements using highly sensitised calibration quartz. A comparison of D e values determined using a Lexsyg reader with single grain D e-values, which have been previously obtained on a Risø reader, is also presented. The results imply that the Lexsyg reader is a highly reliable measurement device with high reproducibility, yielding instrumental uncertainties of around 0.3%.

Language: English
Page range: 327 - 333
Published on: Oct 1, 2014
Published by: Sciendo
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2014 Johanna Lomax, Sebastian Kreutzer, Markus Fuchs, published by Sciendo