Three consecutive failures, one uncertain verdict: A simulation-based reassessment of accreditation-restriction rules in clinical chemistry Cover

Three consecutive failures, one uncertain verdict: A simulation-based reassessment of accreditation-restriction rules in clinical chemistry

Open Access
|Nov 2026
Read online

Abstract

Background

Proficiency testing (PT) is central to laboratory quality assurance; however, repeated unsatisfactory results may reflect both analytical performance and PT-scheme architecture. We evaluated whether three consecutive unsatisfactory results reliably identify deficient analytical performance in serum creatinine testing.

Methods

Monte Carlo simulation modelled three-round trajectories in an ISO 13528-based PT scheme. A total of 206 prespecified scenarios generated 1,446 analytical conditions varying laboratory performance, peer-group definition and size, heterogeneity, consensus construction, between-round persistence and σPT derivation. More than 875 million three-round trajectories were simulated. False-trigger and detection probabilities were evaluated for competent profiles and deficiencies caused by systematic bias, increased analytical imprecision or both.

Results

Under the reference condition, false-trigger probabilities were 0-0.03% for competent profiles, reaching 2.81% under the most adverse configuration examined. Detection probabilities were 20.47% for systematic bias, 0.38% for increased analytical imprecision and 16.24% for combined deficiency. With analytical performance held constant, changes in peer-group definition, focal-laboratory contribution to consensus or σPT derivation altered trigger probabilities by approximately 1.3- to 1.9-fold. Full between-round persistence increased false-trigger probability by approximately 24% relative to independence. Distributional sensitivity analysis produced a maximum absolute difference of 0.69 percentage points.

Conclusions

Three consecutive unsatisfactory PT results are not an invariant indicator of deficient analytical performance. Their probability depends on both the mechanism of deficiency and PT-scheme architecture. Such outcomes should therefore trigger investigation rather than independently establish analytical deficiency or its cause.

DOI: https://doi.org/10.2478/rrlm-2026-0032 | Journal eISSN: 2284-5623 | Journal ISSN: 1841-6624 (formerly 2284-5623)
Language: English
Page range: 297 - 306
Submitted on: Sep 3, 2026
Accepted on: Oct 2, 2026
Published online: Nov 12, 2026
Published in print: Oct 1, 2026
Published by: Romanian Association of Laboratory Medicine
In partnership with: Paradigm Publishing Services

© 2026 Radu Ilinca, Dan Adrian Luțescu, Ionela Ganea, Daniela Cristina Dimitriu, published by Romanian Association of Laboratory Medicine
This work is licensed under the Creative Commons Attribution 4.0 License.