Abstract

Aim

There are many preanalytical variables affecting routine coagulation tests. Increased hematocrit (Htc) levels are one of these variables. However, no study has been conducted to determine the effect of low Htc values on coagulation tests. Therefore, in this study, we aimed to evaluate whether low Htc values affect coagulation tests besides high Htc values.

Methods

Standard human plasma was injected into coagulation tubes containing 3.2% sodium citrate to reflect hematocrit rates of 5% to 75% and prothrombin time (PT), active partial thromboplastin time (aPTT), thrombin time (TT) and fibrinogen measurements were performed.

Results

Three groups were formed according to Htc levels: A (5-25%), B (30-50%) and C (55-75%). PT (s) were found 13.7±0.30 in group A, 14.88±0.57 s in group B, 20.16 ±4.66 s in group C respectively. aPTT (s) results were 35.79±1.39 s in group A, 42.48 ± 3.51 s in group B and 76.47 ± 31.55 in group C. TT (s) results were found to be 26.42 ± 0.77 s (group A), 28.24±1.17 s (group B) and 32.02±2.60 (group C). Fibrinogen levels (g/L) were measured as 2.30 ± 0,05, 2.21 ± 0,07 and 1.90 ± 0.20 in groups A, B, C, respectively. For all measured parameters, group A reflecting low Htc was significantly (P < 0 ,0001) different from the other groups.

Conclusion

Previous studies have reported that high Htc (> 55%) levels affected routine coagulation tests. In our study, low Htc (5-25%) values were also shown to cause errors in the test results.

DOI: https://doi.org/10.2478/rrlm-2023-0014 | Journal eISSN: 2284-5623 | Journal ISSN: 1841-6624
Language: English
Page range: 83 - 90
Submitted on: Dec 27, 2022
Accepted on: Apr 17, 2023
Published on: May 4, 2023
Published by: Romanian Association of Laboratory Medicine
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2023 Ikbal Ozen Kucukcetin, Cahit Nacitarhan, Sebahat Ozdem, published by Romanian Association of Laboratory Medicine
This work is licensed under the Creative Commons Attribution 4.0 License.