Exploring the effectiveness of CARE Dose 4D and CARE kV Automatic Exposure Control Systems in CT: A Slim and Obese Phantom Study Cover

Exploring the effectiveness of CARE Dose 4D and CARE kV Automatic Exposure Control Systems in CT: A Slim and Obese Phantom Study

Open Access
|Sep 2026
Read online

Abstract

Introduction

Radiology strives to minimise radiation exposure while maintaining diagnostic image quality. AEC systems achieve this by optimising radiation doses across different patients and anatomies. The purpose of our study was to evaluate the performance of AEC using Siemens CARE Dose4D and CARE kV options for both slim and obese adult phantoms. We focused on assessing how different parameter settings affect radiation dose and image quality.

Material and methods

Examinations were performed on a dual-source CT scanner using a slim and obese anthropomorphic phantom and a Catphan phantom. Parameters such as mAs modulation strength, reference kV values, quality reference mAs values, dose optimisation settings and CARE kV option were adjusted. The collected data included mA modulation, CTDIvol, effective mAs, quality reference mAs, reference mAs, kV values, and the relative standard deviation of HU values.

Results

The quality reference mAs serves as the baseline target for image quality and is influenced by the CARE kV mode, thereby designating the reference mAs (‘off ’ matches the baseline exactly; ‘on’ scales with the reference kV; ‘semi’ scales inversely with the reference kV). The effective mAs represents the final value influenced by all forms of modulation. For the slim phantom, the effective mAs remained lower than the reference mAs, while for the obese phantom, it was higher. The adjusted reference mAs values appear independent of phantom size. mA modulation curves reflect the link between CTDIvol and AEC. Regardless of the phantom’s body type, CTDIvol decreased/increased when dose-saving optimiser levels, quality mAsref, and kVref were changed. Higher modulation strength lowers the dose for slim phantoms, but raises it for obese phantoms. Slim phantoms produce a lower relative standard deviation than obese ones.

Conclusions

Radiologists and medical physicists must understand the mechanics of the AEC system in CT scanners. This knowledge is essential for ensuring consistent image quality and minimising patient exposure.

DOI: https://doi.org/10.2478/pjmpe-2026-0021 | Journal eISSN: 1898-0309 (formerly 1425-4689) | Journal ISSN: 1425-4689
Language: English
Page range: 193 - 205
Submitted on: Jun 12, 2026
Accepted on: Sep 1, 2026
Published online: Sep 30, 2026
Published in print: Sep 1, 2026
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services

© 2026 Joanna Kidoń-Szołtysek, Aleksandra Klimas, Marcin Dybek, Leszek Wojciuch, Aneta Kawa-Iwanicka, Krzysztof S. Gołba, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.