Have a personal or library account? Click to login
Establishment of national diagnostic dose reference levels (DRLs) for routine computed tomography examinations in Jordan Cover

Establishment of national diagnostic dose reference levels (DRLs) for routine computed tomography examinations in Jordan

Open Access
|Feb 2023

Abstract

Background: Dose reference levels (DRLs) are used as indicators as well as guidance for dose optimization and to ensure justification of appropriate dose for a given clinical indication. The main aims of this study were to establish local DRLs for each CT imaging protocol as a reference point to evaluate the radiation dose indices and to compare our DRLs with those established in other countries and against the internationally reported guidelines.

Materials and methods: 2000 CT dose reports of different adult imaging protocols from January 2021 until April 2022 were collected retrospectively at different hospitals in Jordan. Data were collected from CT scans that were performed using different types and models of CT scanners and included four adult non-enhanced, helical CT imaging protocols; Head, Chest, Abdomen-Pelvis, and Chest-Abdomen-Pelvis.

Results: The average doses of CTDIvol, DLP, and effective dose were (65.11 mGy, 1232.71 mGy·cm, 2.83 mSv) for the head scan, (16.6 mGy, 586.6 mGy·cm, 8.21 mSv) for the chest scan, (17.91 mGy, 929.9 mGy·cm, 13.9 mSv) for the abdomen-pelvis scan, and (19.3 mGy, 1152 mGy·cm, 17.25 mSv) for the chest-abdomen-pelvis scan. In comparison with results from different international studies, DLP values measured in the present study were lower for the chest-abdomen-pelvis and abdomen-pelvis CT scans, and higher for the head CT and chest CT scans.

Conclusions: It is very important that each country establishes its own DRLs and compares them with those reported by other countries, especially the developed ones. It is also important that these levels are regularly updated.

DOI: https://doi.org/10.2478/pjmpe-2023-0003 | Journal eISSN: 1898-0309 | Journal ISSN: 1425-4689
Language: English
Page range: 26 - 34
Submitted on: Oct 13, 2022
Accepted on: Jan 10, 2023
Published on: Feb 1, 2023
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2023 Khaldoon Radaideh, Ali Al-Radaideh, Ramzun Maizan Ramli, Abdallah Saleh, Rasha Alshayeb, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.