Have a personal or library account? Click to login
Calibration of Gafchromic XR-RV3 film under interventional radiology conditions Cover

Calibration of Gafchromic XR-RV3 film under interventional radiology conditions

Open Access
|Jul 2021

Abstract

Introduction: The purpose of the study was the calibration of Gafchromic films in clinical interventional radiology conditions and the assessment of the influence of dose range, the shape of the fitting curve, and its practical application. The aim of the work was to show how practically perform calibration in a wide range of doses.

Material and methods: Gafchromic XR–RV3 films were included in the study. The calibration was performed for A and B film series separately. Doses from the range of 0 – 8 Gy were used. Film dosimeters were read out in reflective mode with a commercial flatbed scanner.

Results: Among various degrees of a polynomial function, the best fit, which fulfilled the chosen criterion of 95% agreement between measured and reconstructed doses and simple equation criterion, was observed for third-degree polynomial. The fitting curve where the dose is the function of optical density (logMPV) was demonstrated to be more precise than the fitting curve based on MPV only. To minimize the difference between dose absorbed by the film and dose reconstructed from the fitting curve below 5% it is necessary to divide the calibration range of 0 – 8 Gy into two subranges for use in interventional radiology. This difference was set at a maximum level of 3.8% and 1.9% for the lowand high-dose range, respectively. Each series of films may have a slightly different calibration curve, especially for the low dose range. A deviation of up to 36% between two batches of Gafchromic film was observed.

Conclusions: For the third-degree polynomial fitting function (one of the recommended in the literature) calibration should be done into low and high dose ranges and for each batch separately. A systematic error higher than 20% could be introduced when the fitting curve from one film batch is applied to the other film batch.

DOI: https://doi.org/10.2478/pjmpe-2021-0020 | Journal eISSN: 1898-0309 | Journal ISSN: 1425-4689
Language: English
Page range: 165 - 173
Published on: Jul 1, 2021
Published by: Polish Society of Medical Physics
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2021 Joanna Kidoń, Kinga Polaczek-Grelik, Krzysztof Gołba, Wojciech Wojakowski, Andrzej Ochała, published by Polish Society of Medical Physics
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 License.