Detection of physiological concentrations of GABA using dielectric spectroscopy - A pilot study Cover

Detection of physiological concentrations of GABA using dielectric spectroscopy - A pilot study

Open Access
|Dec 2023

Figures & Tables

Figure 1:

Raw dielectric constant spectra as a function of frequency for 6 GABA concentrations investigated. 100 spectra for each of the concentrations. From [13], with permission.
Raw dielectric constant spectra as a function of frequency for 6 GABA concentrations investigated. 100 spectra for each of the concentrations. From [13], with permission.

Figure 2:

Water contribution subtracted dielectric constant spectra, as a function of frequency for 6 GABA concentrations. The average of 100 measurements of the water dielectric constant spectrum was subtracted from each of the GABA solution measurements. From [13], with permission.
Water contribution subtracted dielectric constant spectra, as a function of frequency for 6 GABA concentrations. The average of 100 measurements of the water dielectric constant spectrum was subtracted from each of the GABA solution measurements. From [13], with permission.

Figure 3:

Principal component analysis on six different GABA concentrations. The X-axis shows the first component which covers 81.8% of the variance in the dataset, the Y-axis shows the second component which covers 10.4% of the variance in the dataset. From [13], with permission.
Principal component analysis on six different GABA concentrations. The X-axis shows the first component which covers 81.8% of the variance in the dataset, the Y-axis shows the second component which covers 10.4% of the variance in the dataset. From [13], with permission.

LSTM model results on a dataset consisting of six different GABA concentrations_ “Raw data” means unprocessed dielectric constant data obtained from the GABA solution measurements_ “Subtracted data” implies that the experimental dielectric constant of water (data averaged over 100 measurements) was subtracted from each of the dielectric constant data for the GABA solutions_ From [13], with permission_

Test AccuracyTest Loss
Raw data0.901640.39048
Subtracted data0.991670.01064

Performance of the LSTM neural network_ Raw ϵ′ data is the measured dielectric constant data of GABA solutions, whereas the subtracted ϵ′ data is the difference between the measured dielectric constant data of the GABA solutions and the average dielectric constant data of deionized water_ From [13], with permission_

Raw ϵ′ dataSubtracted ϵ′ data
c [μm]Test dataset AccuracyTest dataset LossTest dataset AccuracyTest dataset Loss
5 vs 101.00.006091.00.00007
10 vs 201.00.004491.00.00004
20 vs 501.00.001391.00.00009
50 vs 1001.00.064001.00.00270
100 vs 2001.00.083031.00.00204
Language: English
Page range: 47 - 52
Submitted on: Nov 15, 2023
Published on: Dec 31, 2023
Published by: University of Oslo
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2023 Jie Hou, Runar Strand-Amundsen, Ørjan G. Martinsen, published by University of Oslo
This work is licensed under the Creative Commons Attribution 4.0 License.