Calibration of a Measuring Device to Determine Specific Saturation Magnetization by A Special Coil Cover

Calibration of a Measuring Device to Determine Specific Saturation Magnetization by A Special Coil

By: Michal Ulvr  
Open Access
|Aug 2014

Abstract

Specific saturation magnetization σS is an important parameter of ferromagnetic materials. It is in principle independent of the structure and shape of the sample. There are two ways to calibrate devices for determining σS: the standard method, using a calibrated magnetic moment standard, or an alternative method, using a special moment coil. This paper presents a calibration method used at Czech Metrology Institute for calibration of Koerzimat 1.096 using a special moment coil. Special attention is given to a description of the special moment coils used for Koerzimat 1.096 calibration, and to analysis of sources of uncertainty. Calibration expanded uncertainty of 0.4 up to 0.6 % can be achieved using this method.

DOI: https://doi.org/10.2478/jee-2014-0041 | Journal eISSN: 1339-309X | Journal ISSN: 1335-3632
Language: English
Page range: 262 - 264
Submitted on: Jan 7, 2014
Published on: Aug 21, 2014
Published by: Slovak University of Technology in Bratislava
In partnership with: Paradigm Publishing Services
Publication frequency: 6 issues per year

© 2014 Michal Ulvr, published by Slovak University of Technology in Bratislava