Fig. 1.

Flowchart depicting the screening and enrolment process within this study. PIVC, peripheral intravascular catheter.
Flowchart depicting the screening and enrolment process within this study. PIVC, peripheral intravascular catheter.

Structure of newly developed and existing needles

Catheter gauge sizeTypes of needles (newly developed and existing needles)Geometric design of the needleBevel length, mmBevel angle, degrees
20GSurshield SurfloII with the thin-tipped short bevel needle named 3D-Shin (newly developed needle)2.222
Surshield SurfloII (existing needle)2.228
22GSurshield SurfloII with the thin-tipped short bevel needle named 3D-Shin (newly developed needle)1.922
Surshield SurfloII (existing needle)1.926

Result of multiple regression analysis for newly developed needles and the number of procedures†

VariablesUnivariate analysis Regression coefficient (95% CI)p-valueMultivariable analysis Regression coefficient (95% CI)p-value
Types of needles
Existing needleref-ref-
Newly developed needle0.00021 [(−0.059)–(0.063)]0.95−0.0042 [(−0.065)–(0.056)]0.89

Patient characteristics during emergency room arrival

VariablesOverall (n = 522)Newly developed needles (n = 261)Exiting needles (n = 261)p-value
Age, median (IQR), years73 (57–82)73 (56–82)73 (57–81)0.94
Male sex (n, %)297 (56.9)142 (54.4)155 (59.4)0.29
Body height, median (IQR), cm162 (155–168)161 (155–168)162 (155–169)0.49
Body weight, median (IQR), kg57 (50–67)57 (50–66)57 (50–68)0.54
BMI, median (IQR), kg/m222.0 (19.5–24.2)21.8 (19.1–24.3)22.0 (19.5–24.2)0.66
Charlson Comorbidity Index, median (IQR)1 (0–2)1 (0–2)1 (0–2)0.32
SOFA score, median (IQR)0 (0–1)0 (0–1.8)0 (0–1)0.93
Hypertension (n, %)209 (40.0)103 (39.5)106 (40.6)0.86
Dyslipidemia (n, %)85 (16.3)36 (13.8)49 (18.8)0.16

Insertion site (n, %)
Hand13 (2.5)10 (3.8)3 (1.1)0.05
Forearm385 (73.8)185 (70.9)200 (76.6)0.44
Elbow113 (21.7)58 (22.2)55 (21.1)0.78
Upper arm10 (1.9)7 (2.7)3 (1.1)0.21
Lower leg1 (0.2)1 (0.4)0 (0)-
Visibility of veins (n, %)462 (88.5)230 (88.1)232 (88.9)0.89
Palpability of veins (n %)490 (93.9)243 (93.1)247 (94.6)0.59
Mobility of veins (n, %)249 (47.7)126 (48.3)123 (47.1)0.86

Inserted by (n, %)
Residents5 (1.0)3 (1.1)2 (0.8)0.65
Physicians excluding residents37 (7.1)19 (7.3)18 (6.9)0.87
Individuals with < 5 years of nursing experience82 (15.7)42 (16.1)40 (15.3)0.83
Individuals with ≥ 6 years of nursing experience398 (76.3)197 (75.7)201 (77.0)0.84
Catheter gauge (n, %)
20G330 (63.2)161 (61.7)169 (64.8)0.53
22G192 (36.8)100 (38.3)92 (35.2)0.56
Major axis of veins, median (IQR), mm3.0 (2.4–4.0)3.0 (2.2–4.0)3.2 (2.6–4.0)0.02
Major axis of vessels/diameter of PIVCs, median (IQR)3.1 (2.3–3.8)3.0 (2.3–3.7)3.2 (2.6–3.9)0.04
Movements of patients (n, %)27 (5.2)13 (5.0)14 (5.4)1.0
Failure of initial insertion (n, %)81 (15.5)39 (14.9)42 (16.1)0.81
Number of procedures performed until the success of insertion††, median (IQR)1 (1–1)1 (1–1)1 (1–1)0.86

Result of logistic regression analysis for newly developed needles and failure of initial PIVC insertion

VariablesUnivariate analysis OR (95% CI)p-valueMultivariable analysis OR (95% CI)p-value
Types of needles
Existing needleref-ref-
Newly developed needle0.92 (0.57–1.47)0.720.79 (0.48–1.31)0.36
DOI: https://doi.org/10.2478/jccm-2024-0025 | Journal eISSN: 2393-1817 | Journal ISSN: 2393-1809
Language: English
Page range: 213 - 221
Submitted on: Mar 13, 2024
Accepted on: Jun 2, 2024
Published on: Jul 31, 2024
Published by: University of Medicine, Pharmacy, Science and Technology of Targu Mures
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2024 Yuki Kishihara, Hideto Yasuda, Masahiro Kashiura, Takatoshi Oishi, Yutaro Shinzato, Takashi Moriya, published by University of Medicine, Pharmacy, Science and Technology of Targu Mures
This work is licensed under the Creative Commons Attribution 4.0 License.