Abstract

Introduction: Sterilisation using peracetic acid (PAA) has been advocated for orthodontic elastic bands. However, cane-loaded elastomeric ligatures can also become contaminated during processing, packaging, and manipulation before placement in the oral cavity and are therefore susceptible, and possible causes, of cross-contamination.

Aim: To test the hypothesis that 0.25% peracetic acid (PAA), following the sterilisation of elastomers, influences the cytotoxicity of elastomeric ligatures on L929 cell lines.

Materials and methods: Four hundred and eighty silver elastomeric ligatures were divided into 4 groups of 120 ligatures to produce, Group TP (latex natural, bulk pack, TP Orthodontics), Group M1 (Polyurethane, bulk pack, Morelli), Group M2 (Polyurethane, cane-loaded, Morelli) and Group U (Polyurethane, cane-loaded, Uniden). Of the 120 ligatures in each group, 100 were sterilised in 0.25% PAA at time intervals (N = 20) of 1 hour, 2 hours, 3 hours, 4 hours and 5 hours. The 20 remaining elastomeric ligatures in each group were not sterilised and served as controls. Cytotoxicity was assessed using L929 cell lines and a dye-uptake method. Analysis of variance (ANOVA), followed by the Tukey post hoc test (p < 0.05) determined statistical relevance.

Results: There was a significant difference between TP, Morelli and Uniden elastomerics (p < 0.05), but no difference between the two types of Morelli elastomerics at the 1 hour time interval. In addition, there was a significant difference between Group CC and the other groups assessed, except between Groups CC and TP at the 1 hour time interval. The non-sterilised elastomeric ligatures showed similar cell viability to that observed after 1 hour of standard sterilisation.

Conclusion: PAA did not significantly influence the cytotoxicity of elastomeric ligatures after a sterilisation time of 1 hour and is therefore recommended for clinical use.

DOI: https://doi.org/10.2478/aoj-2013-0015 | Journal eISSN: 2207-7480 | Journal ISSN: 2207-7472
Language: English
Page range: 139 - 144
Submitted on: Jan 1, 2012
Accepted on: May 1, 2013
Published on: Aug 1, 2023
Published by: Australian Society of Orthodontists Inc.
In partnership with: Paradigm Publishing Services
Publication frequency: 1 issue per year

© 2023 Matheus Melo Pithon, Rogerio Lacerda dos Santos, Renata Lima Pasini Judice, Paulo Sergio de Assuncao, Luciana Restle, published by Australian Society of Orthodontists Inc.