Have a personal or library account? Click to login
Resection of the Tooth Apex with Diode Laser Cover

Resection of the Tooth Apex with Diode Laser

Open Access
|Nov 2014

Abstract

An “in vitro” experimental study has been carried out on 70 extracted teeth. A laser resection of the root apex has been carried out with diode laser beam with a wavelength of - 810 ± 10 nm. Sequentially a radiation with increasing power has been applied, as follows: 1,3 W, 2W, 3W, 4W, 5W, 6W, 7W, in electro surgery mode. Successful resection of the tooth apex has been performed at: 3W; 4W; 5W; 6W and 7W power. It was established that when laser resected the tooth apex carbonizes.

DOI: https://doi.org/10.2478/amb-2014-0006 | Journal eISSN: 2719-5384 | Journal ISSN: 0324-1750
Language: English
Page range: 45 - 49
Published on: Nov 5, 2014
Published by: Sofia Medical University
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2014 Tz. Uzunov, P. Uzunova, T. Uzunov, R. Grozdanova, published by Sofia Medical University
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.