Have a personal or library account? Click to login
Effect of photon energy spectrum on dosimetric parameters of brachytherapy sources Cover

Effect of photon energy spectrum on dosimetric parameters of brachytherapy sources

Open Access
|Mar 2016

Abstract

Aim

The aim of this study is to quantify the influence of the photon energy spectrum of brachytherapy sources on task group No. 43 (TG-43) dosimetric parameters.

Background

Different photon spectra are used for a specific radionuclide in Monte Carlo simulations of brachytherapy sources.

Materials and methods

MCNPX code was used to simulate 125I, 103Pd, 169Yb, and 192Ir brachytherapy sources. Air kerma strength per activity, dose rate constant, radial dose function, and two dimensional (2D) anisotropy functions were calculated and isodose curves were plotted for three different photon energy spectra. The references for photon energy spectra were: published papers, Lawrence Berkeley National Laboratory (LBNL), and National Nuclear Data Center (NNDC). The data calculated by these photon energy spectra were compared.

Results

Dose rate constant values showed a maximum difference of 24.07% for 103Pd source with different photon energy spectra. Radial dose function values based on different spectra were relatively the same. 2D anisotropy function values showed minor differences in most of distances and angles. There was not any detectable difference between the isodose contours.

Conclusions

Dosimetric parameters obtained with different photon spectra were relatively the same, however it is suggested that more accurate and updated photon energy spectra be used in Monte Carlo simulations. This would allow for calculation of reliable dosimetric data for source modeling and calculation in brachytherapy treatment planning systems.

DOI: https://doi.org/10.1515/raon-2016-0019 | Journal eISSN: 1581-3207 | Journal ISSN: 1318-2099
Language: English
Page range: 238 - 246
Submitted on: Nov 2, 2015
|
Accepted on: Jan 29, 2016
|
Published on: Mar 31, 2016
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2016 Mahdi Ghorbani, Mohammad Mehrpouyan, David Davenport, Toktam Ahmadi Moghaddas, published by Association of Radiology and Oncology
This work is licensed under the Creative Commons Attribution-NonCommercial-NoDerivatives 3.0 License.